Non‐invasive vs. invasive beat‐to‐beat monitoring of blood pressure

View Full Paper
AGA GabrielLLL. E. LindbladCAC. Angleryd

Structured PICO

Does non-invasive beat-to-beat blood pressure monitoring accurately trace fast changes compared to invasive intra-arterial radial pressure in patients?

P
Population
8 patients studied at rest, during deep breathing with a fixed rate of 6 breaths min-1, and during an exercise test on an ergometer cycle
I
Intervention
Non-invasive beat-to-beat blood pressure monitoring (Finapres)
C
Comparator
Ipsilateral intra-arterial radial pressure
O
Outcome
Accuracy in tracing fast beat-to-beat changes in blood pressure (correlation coefficient)surrogate

Non-invasive beat-to-beat blood pressure monitoring using the Finapres device is highly correlated with invasive intra-arterial measurements and is well-suited to track relative changes in blood pressure.

Abstract

The present study focuses on the accuracy in tracing fast beat-to-beat changes in blood pressure using a non-invasive technique. The measurements using a commercially available apparatus (Finapres, Ohmeda, USA) were compared to ipsilateral intra-arterial radial pressure. Eight patients were studied at rest, during deep breathing with a fixed rate of 6 breaths min-1, and during an exercise test on an ergometer cycle. A total of 900 systolic pressure values were included for statistical evaluation, covering a pressure range of 86-266 mmHg. On average the systolic correlation coefficient for the whole material was 0.97, with a range of 0.94-0.996. For mean pressure the correlation coefficient was on average 0.97, and for diastolic pressure 0.93. No systematic difference between the non-invasive and the invasive method was found, although for each individual patient a difference between direct and indirect measured blood pressure existed that could be relatively large (systolic pressure: average difference = 0.8 mmHg, SD = 16 mmHg). We found the method easy to handle and consider it excellently suited to track relative changes in blood pressure.