Key result
Optimized programming of implantable cardioverter defibrillators significantly reduces the rate of inappropriate ventricular tachyarrhythmia detection and shocks.
Why the study?
Does optimized programming of implantable cardioverter defibrillators reduce inappropriate therapy?
Does optimized programming of implantable cardioverter defibrillators reduce inappropriate therapy?
Optimized ICD programming, including prolonged detection times and higher rate cutoffs, is crucial for minimizing inappropriate shocks and improving patient quality of life.
Answers are grounded in this paper and cite their sources.
Supports individualized ICD programming consideration; leaves open prospective validation in contemporary cohorts before practice change.
PURPOSE OF REVIEW: With wider indications for implantable cardioverter defibrillator therapy, more patients at lower risk for ventricular tachyarrhythmia receive this treatment. To maintain the ratio of benefit versus side-effects at an acceptable level, the risk of inappropriate implantable cardioverter defibrillator therapy has to be minimized. RECENT FINDINGS: Implantable cardioverter defibrillators require the activation of enhanced detection criteria. These can avoid inappropriate therapy of sinus tachycardia (gradual onset) and atrial fibrillation (irregular rate) while other regular supraventricular tachycardias may be misclassified even with combinations of criteria (QRS morphology, abrupt onset, regular rate). Carefully programmed, dual-chamber implantable cardioverter defibrillators provide better ventricular tachyarrhythmia/supraventricular tachycardias discrimination. Key issues are long tachycardia detection (18 cycles or more), deactivation or restrictive programming of safety therapy despite supraventricular tachycardia classification, and restriction of shock therapy to high tachycardia rates (>or=250 bpm). Further developments are necessary to reduce the incidence of inappropriate therapy due to lead failure that is more frequent in physically active patients. SUMMARY: With optimized programming, the rate of inappropriate ventricular tachyarrhythmia detection is significantly reduced. Particularly the prevention of inappropriate shocks has important implication for the quality of life and acceptance of implantable cardioverter defibrillator treatment.
No takes yet. Share an insight, caveat, or question.
Carsten W. Israel (2008) conducted a review in Implantable cardioverter defibrillator therapy. Optimized ICD programming was evaluated on Inappropriate ventricular tachyarrhythmia detection and shocks. Optimized programming of implantable cardioverter defibrillators significantly reduces the rate of inappropriate ventricular tachyarrhythmia detection and shocks.
Synapse has enriched 10 related papers. Consider them for comparative context: