Key result
CT calcium volume ≥337 mm³ outside the virtual valve predicts significant perivalvular leak with Sapien 3.
Why the study?
To evaluate the predictive ability of quantifying aortic valve calcification using a virtual valve on CT imaging for perivalvular leak after Sapien 3 transcatheter heart valve implantation.
Does quantifying aortic valve calcification using a virtual valve on CT imaging predict significant perivalvular leak in patients undergoing Sapien 3 implantation?
Observational (n=126)
Does quantifying aortic valve calcification using a virtual valve on CT imaging predict significant perivalvular leak in patients undergoing Sapien 3 implantation?
Effect estimate: AUC 0.82
p-value: p=<0.0001
Quantifying calcium volume outside a virtual valve on pre-procedural CT imaging accurately predicts significant perivalvular leak after Sapien 3 TAVR.
BACKGROUND: This study aimed to evaluate the predictive ability of quantifying aortic valve calcification using a virtual valve (VV) on computed tomography (CT) imaging for perivalvular leak (PVL) following Sapien 3 transcatheter heart valve implantation. METHODS: A total of 126 patients who underwent Sapien 3 implantation were retrospectively analyzed. Patients with left ventricular outflow tract calcification or bicuspid valves were excluded. Multiplanar CT reconstructions of the aortic root with the VV in position were utilized to measure calcium volumes located outside the VV, inside the VV, and across the entire valve leaflet. PVL was quantitatively assessed via transthoracic echocardiography using the formula: %PVL = (jet length along valve curvature/total perimeter) × 100. Significant PVL was defined as %PVL > 10%. RESULTS: Calcium volumes outside the VV, inside the VV, and in the entire valve leaflets were significantly correlated with %PVL (r = 0.58, p < 0.0001; r = 0.38, p < 0.0001; r = 0.50, p < 0.0001, respectively). Receiver operating characteristic (ROC) curve analysis identified optimal cutoff values for predicting significant PVL: calcium volume outside the VV ≥ 336.9 mm³ (area under the curve [AUC] 0.82, sensitivity 100%, specificity 59.3%), inside the VV ≥ 2502.1 mm³ (AUC 0.68, sensitivity 45.0%, specificity 90.7%), and in the entire valve leaflets ≥ 3268.2 mm³ (AUC 0.74, sensitivity 45.0%, specificity 94.2%). Among these, calcium volume outside the VV demonstrated the highest AUC. CONCLUSIONS: Quantifying calcium volume using VV on CT imaging is a valuable tool for accurately predicting significant PVL after Sapien 3 implantation.
No takes yet. Share an insight, caveat, or question.
Yabuta et al. (2025) conducted an observational in Sapien 3 transcatheter heart valve implantation (n=126). Virtual valve-based computed tomography imaging for calcium quantification was evaluated on Significant perivalvular leak (%PVL > 10%) (AUC 0.82, p=<0.0001). Calcium volume outside the virtual valve on CT imaging ≥ 336.9 mm³ optimally predicted significant perivalvular leak after Sapien 3 implantation (AUC 0.82, sensitivity 100%, specificity 59.3%).