August 1, 2012

Model-based error analysis of the oscillometric fixed-ratio blood pressure measurement method

View Full Paper
Ask AI
Bookmark
Share

Key result

A mathematical model demonstrated that increased arterial stiffness and pulse pressure cause the fixed-ratio method to predict higher systolic and lower diastolic blood pressures.

Population

Mathematical model of oscillometry

Design

Other

Authors

JLJiankun LiuMitsubishi Heavy Industries (Germany)JHJin‐Oh HahnPreventive CardiologyRMRamakrishna MukkamalaPreventive Cardiology

Discussion

Loading...

Member takes

Implication

May bias fixed-ratio oscillometric readings in stiff arteries; leaves open whether algorithms require stiffness adjustment.

Structured PICO

P
Population
Mathematical model of oscillometry
I
Intervention
Parametric sensitivity analysis of the oscillometric fixed-ratio blood pressure measurement method
O
Outcome
Accuracy and error of the fixed-ratio blood pressure measurement method

Theoretical modeling indicates that arterial stiffness and pulse pressure are major factors that can cause significant errors in oscillometric fixed-ratio blood pressure measurements.

Cite This Study

Liu et al. (2012) studied Blood pressure measurement accuracy. Oscillometric fixed-ratio blood pressure measurement method was evaluated on Measurement error in systolic and diastolic blood pressures. A mathematical model demonstrated that increased arterial stiffness and pulse pressure cause the fixed-ratio method to predict higher systolic and lower diastolic blood pressures.

synapsesocial.com/papers/6a13fe733f92ec2dd7599874https://doi.org/10.1109/embc.2012.6346011
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Errors of oscillometric blood pressure measurement as predicted by simulation2011 · 33 citations
  2. 2The Meaning of the Point of Maximum Oscillations in Cuff Pressure in the Indirect Measurement of Blood Pressure—Part II1980 · 223 citations
  3. 3Accuracy of the Omron HEM-705 CP for blood pressure measurement in large epidemiologic studies.2011 · 63 citations
  4. 4Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension protocol2003 · 22 citations
  5. 5Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M72009 · 48 citations