Key result
Transcatheter aortic valve replacement using the Lotus valve resulted in 95% VARC-2 hemodynamic success and a 0.4% rate of moderate or greater paravalvular leak at 1-year.
Why the study?
The Lotus aortic valve system is designed with a seal to minimize PVL, motivating evaluation of hemodynamic performance, PVL, and ventricular function up to 1 year after TAVR in severe aortic stenosis.
Does TAVR using the Lotus THV improve hemodynamic performance and minimize paravalvular leak up to 1-year in patients with severe aortic stenosis?
Observational (n=550)
Yes
Does TAVR using the Lotus THV improve hemodynamic performance and minimize paravalvular leak up to 1-year in patients with severe aortic stenosis?
TAVR with the Lotus valve demonstrates persistently favorable hemodynamic performance and a low rate of paravalvular leak through 1 year in patients with severe aortic stenosis.
Supports Lotus TAVR hemodynamics in severe AS; leaves open randomized comparisons to contemporary valves.
BackgroundThe Lotus aortic valve system is a transcatheter heart valve (THV) designed with a seal to minimize paravalvular leak (PVL). We evaluated hemodynamic performance, PVL and ventricular function up to 1-year after transcatheter aortic valve replacement (TAVR) using the Lotus THV in patients with severe aortic stenosis.MethodsRESPOND was a prospective, single-arm, multicenter registry evaluating outcomes after Lotus TAVR in routine clinical practice. Core Laboratory adjudicated echocardiograms at baseline, discharge and 1-year were analyzed in a subgroup (N = 550) of patients with paired transthoracic echocardiography (TTE) at the 3 time points.ResultsMean effective aortic valve orifice area (EOA) in cm2 increased from 0.74 ± 0.25 at baseline to 1.83 ± 0.46 at discharge and to 1.78 ± 0.45 at 1-year. Mean transvalvular gradient in mmHg was reduced from 38.7 ± 15.2 at baseline to 10.9 ± 4.2 at discharge and to 10.8 ± 5.1 at 1-year. VARC-2 hemodynamic success was achieved in 92% and 95% of patients at discharge and 1-year, respectively. Mild or more PVL was seen in 7.3% and 5.5% and moderate or more PVL was seen in 0.6% and 0.4% of the patients at discharge and 1-year, respectively. Stroke volume index (mL/m2) increased from 35 ± 10 at baseline to 40 ± 11 at discharge and to 41 ± 11 at 1-year (p < 0.001).ConclusionsThis RESPOND sub-study demonstrates persistently favorable hemodynamic performance and a low rate of PVL in unselected patients through 1-year in the largest Lotus-treated population to date.(Repositionable Lotus Valve System–Post Market Evaluation of Real World Clinical Outcomes [RESPOND]; NCT02031302).
No takes yet. Share an insight, caveat, or question.
Soliman et al. (2019) conducted an observational in Severe aortic stenosis (n=550). Lotus transcatheter heart valve (THV) was evaluated on VARC-2 hemodynamic success at 1-year. Transcatheter aortic valve replacement using the Lotus valve resulted in 95% VARC-2 hemodynamic success and a 0.4% rate of moderate or greater paravalvular leak at 1-year.