Key result
No commercially available cardiac troponin assay was able to achieve the recommended 10% coefficient of variation at the 99th percentile reference limit defined by the manufacturer.
Why the study?
Do commercially available cardiac troponin assays achieve a 10% coefficient of variation at the 99th percentile reference limit?
Do commercially available cardiac troponin assays achieve a 10% coefficient of variation at the 99th percentile reference limit?
At the time of the study, no commercially available cardiac troponin assay could achieve the recommended 10% imprecision (CV) at the 99th percentile reference limit for diagnosing myocardial infarction.
No takes yet. Share an insight, caveat, or question.
Current assays undermine reliable MI diagnosis at manufacturer 99th percentiles; challenges ESC/ACC imprecision standards and leaves open revised criteria.
Panteghini et al. (2004) studied Myocardial Infarction. Commercially available cardiac troponin assays was evaluated on Lowest concentration associated with a 10% CV. No commercially available cardiac troponin assay was able to achieve the recommended 10% coefficient of variation at the 99th percentile reference limit defined by the manufacturer.