Key result
Doppler pressure gradients showed good agreement with catheter measurements in 3D-printed mitral valves, despite rapid-prototyped valve dimension mean relative errors ranging from 6.2% to 35%.
Why the study?
Does Doppler-derived transmitral pressure gradient accurately reflect catheter-measured gradients in 3D models of mitral valve calcification?
Does Doppler-derived transmitral pressure gradient accurately reflect catheter-measured gradients in 3D models of mitral valve calcification?
In an in vitro model of mitral annular calcification, Doppler-derived pressure gradients showed good agreement with catheter measurements, supporting its clinical use for estimating transmitral gradients.
Answers are grounded in this paper and cite their sources.
Agreement in 3D-printed mitral models supports in vitro validation; leaves open clinical translation pending in vivo studies.
Mitral annular calcification (MAC) involves calcium deposition in the fibrous annulus supporting the mitral valve (MV). When calcification extends onto the leaflets, valve opening can be restricted. The influence of MAC MV geometry on Doppler gradients is unknown. This study describes a novel methodology to rapid-prototype subject-specific MAC MVs. Replicated valves were used to assess the effects of distorted annular-leaflet geometry on Doppler-derived, transmitral gradients in comparison to direct pressure measurements and to determine if transmitral gradients vary according to measurement location. Three-dimensional echocardiography data sets were selected for two MAC MVs and one healthy MV. These MVs were segmented and rapid prototyped in their middiastolic configuration for in vitro testing. The effects of MV geometry, measurement modality, and measurement location on transmitral pressure gradient were assessed by Doppler and catheter at three locations along the MV's intercommissural axis. When comparing dimensions of the rapid-prototyped valves to the subject echocardiography data sets, mean relative errors ranged from 6.2% to 35%. For the evaluated MVs, Doppler pressure gradients exhibited good agreement with catheter-measured gradients at a variety of flow rates, though with slight systematic overestimation in the recreated MAC valves. For all of the tested MVs, measuring the transmitral pressure gradient at differing valve orifice positions had minimal impact on observed gradients. Upon the testing of additional normal and calcific MVs, these data may contribute to an improved clinical understanding of MAC-related mitral stenosis. Moreover, they provide the ability to statistically evaluate between measurement locations, flow rates, and valve geometries for Doppler-derived pressure gradients. Determining these end points will contribute to greater clinical understanding for the diagnosis MAC patients and understanding the use and application of Doppler echocardiography to estimate transmitral pressure gradients.
No takes yet. Share an insight, caveat, or question.
Herrmann et al. (2013) studied Mitral annular calcification (n=3). Doppler echocardiography vs. Catheter-measured pressure gradients was evaluated on Transmitral pressure gradient. Doppler pressure gradients showed good agreement with catheter measurements in 3D-printed mitral valves, despite rapid-prototyped valve dimension mean relative errors ranging from 6.2% to 35%.
Synapse has enriched 10 related papers. Consider them for comparative context: