Key result
Pentaspline pulsed field ablation achieves 100% acute pulmonary vein isolation, though isolation area regresses acutely.
Why the study?
Pentaspline pulsed field ablation catheter lesions can differ by pose, repetition, and contact, and characterizing lesion morphology may offer insights into suboptimal ablation durability and help optimize ablation strategies.
What are the endocardial atrial lesion characteristics and acute durability following pentaspline pulsed field ablation?
Observational (n=25)
What are the endocardial atrial lesion characteristics and acute durability following pentaspline pulsed field ablation?
Pentaspline pulsed field ablation achieves acute PV isolation reliably, but significant acute regression of the isolation area occurs within 5 minutes, highlighting the need for dose-related footprint assessments to optimize durability.
Answers are grounded in this paper and cite their sources.
Acute PVI with pentaspline PFA shows rapid area regression; leaves open optimal dosing for durable isolation.
Background Pentaspline ablation (31 mm, Farapulse) catheter lesions can differ by pose, repetition, and contact, thus affecting morphology. Objective Lesion morphology may offer insights into suboptimal ablation durability and can help optimize ablation strategies. Methods In 25 patients, 1–2 applications were delivered to pulmonary veins (PV, n = 16), posterior left atrium (LA, n = 6) and LA roof ( n = 3). Pre, 1‐ and 5‐min postablation maps were created. Isolation area was identified by < 0.15 mV. Three patients after cavotricuspid isthmus (CTI) ablation were mapped. Results Despite only 1 or 2 basket pose applications without rotation, all 16/16 PVs (100%) were acutely isolated. Only 3/16 PVs (12.5%) reconnected by 5 min revealing asymmetric, band‐like perivenous lesions. Posterior LA flower applications were irregular, with the smallest mean dimensions of 14.9 and 17.7 mm. Isolation area regressed acutely, more so for single but also for double applications. However, regression occurred only for PV (6.7 ± 6.0 vs. 1.3 ± 1.5 cm 2 ) and not for posterior wall lesions (2.9 [1.9–3.6] vs. 2 [1.65–2.4] cm 2 ). Double roof applications resulted in circular, doughnut‐shaped, or even minimal lesions. Flower pose CTI lesions had median widths of 43.1, 21.3, and 24.9 mm along the line. Conclusion Acute PV isolation always occurs despite minimal dose delivery and is unreliable for lesions assessment. Even double applications have relevant regression, and this occurs more so in PV than LA posterior wall. Despite 2 applications, posterior LA lesions were smaller than the catheter diameter and roof lesions footprints can vary widely. Dose related footprint assessments are needed to help optimize ablation and durability.
No takes yet. Share an insight, caveat, or question.
Banai et al. (2025) conducted an observational in Arrhythmia requiring ablation (n=25). Pentaspline pulsed field ablation (Farapulse) was evaluated on Acute pulmonary vein isolation. Pentaspline pulsed field ablation achieved 100% acute pulmonary vein isolation with 1-2 applications, though isolation area regressed acutely by 5 minutes (6.7 vs 1.3 cm2).
Synapse has enriched 10 related papers. Consider them for comparative context: