Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
August 1, 2008Circulation Cardiovascular InterventionsOpen Access

High Incidence of Intramural Thrombus After Overlapping Paclitaxel-Eluting Stent Implantation

View Full Paper
Ask AI
Bookmark
Share

Key result

Overlapping paclitaxel-eluting stents in a porcine model reduced late lumen loss (0.46 vs 1.30 mm, P=0.01) but caused marked intramural thrombi compared to bare-metal stents.

Why the study?

Does overlapping paclitaxel-eluting stent implantation compared to bare-metal stents affect angioscopic and histopathologic outcomes in pig coronary arteries at 1 month?

Population

Pig coronary arteries (n=9 pigs)

Comparison

Overlapping paclitaxel-eluting stents (PES) vs Overlapping bare-metal stents (BMS)

Design

Preclinical

Follow-up

1 month

Authors

TSToshiro ShinkeInterventional CardiologyJLJinsheng LiChinese PLA General HospitalJCJack P. ChenNorthside Hospital

Discussion

Loading...

Member takes

Implication

May increase intramural thrombosis risk with overlapping paclitaxel-eluting stents; leaves open human applicability and need for clinical trials.

Structured PICO

Does overlapping paclitaxel-eluting stent implantation compared to bare-metal stents affect angioscopic and histopathologic outcomes in pig coronary arteries at 1 month?

P
Population
Pig coronary arteries (n=9 pigs)
I
Intervention
Overlapping paclitaxel-eluting stents (PES)
C
Comparator
Overlapping bare-metal stents (BMS)
O
Outcome
1-month angioscopic and histopathologic sequelae (including late lumen loss, mural thrombi, neointimal thickness, and endothelium-dependent relaxation)surrogate

Main Result

Absolute Event Rate: 0.46% vs 1.3%

p-value: p=0.01

In a porcine model, overlapping paclitaxel-eluting stents reduce neointimal thickness but are associated with marked intramural thrombi and impaired adjacent endothelial function compared to bare-metal stents at 1 month.

Cite This Study

Shinke et al. (2008) studied Coronary stenting (porcine model) (n=9). Overlapping paclitaxel-eluting stents (PES) vs. Bare-metal stents (BMS) was evaluated on Late lumen loss (p=0.01). Overlapping paclitaxel-eluting stents in a porcine model reduced late lumen loss (0.46 vs 1.30 mm, P=0.01) but caused marked intramural thrombi compared to bare-metal stents.

synapsesocial.com/papers/6a154190d73ae7522a4e2447https://doi.org/10.1161/circinterventions.108.770008

Topics

Coronary artery diseaseCoronary CT angiography
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Indication of long-term endothelial dysfunction after sirolimus-eluting stent implantation2005 · 386 citations
  2. 2Neointimal coverage of sirolimus-eluting stents at 6-month follow-up: evaluated by optical coherence tomography2006 · 332 citations
  3. 3Pathological Correlates of Late Drug-Eluting Stent Thrombosis2007 · 1,276 citations
  4. 4Overlapping paclitaxel-eluting stents: Long-term effects in a porcine coronary artery model2007 · 59 citations
  5. 5Safety and Efficacy of Sirolimus- and Paclitaxel-Eluting Coronary Stents2007 · 1,631 citations