Key result
Zero-fluoroscopy LBBP achieves 100% procedural success and narrows QRS width with no complications.
Why the study?
Fluoroscopy during pacemaker implantations causes harmful effects, but there were no prior experiences with zero-fluoroscopy left bundle branch pacing.
Is zero-fluoroscopy left bundle branch pacing feasible and safe in patients requiring pacemaker implantation?
Observational (n=10)
Is zero-fluoroscopy left bundle branch pacing feasible and safe in patients requiring pacemaker implantation?
Zero-fluoroscopy left bundle branch pacing appears feasible and safe, achieving successful implantation and stable pacing parameters without radiation exposure.
Supports feasibility of zero-fluoroscopy LBBP without complications; leaves open need for randomized trials before practice change.
INTRODUCTION: Left bundle branch pacing (LBBP) has emerged in recent years as a new pacing modality, providing patients with a narrower paced QRS than conventional pacing and stable pacing parameters. At the same time, there is a growing concern about the use of fluoroscopy in pacemaker implantations, given its harmful effects on both patients and operators. However, there are no prior experiences of zero-fluoroscopy in LBBP procedure. METHODS: We conducted an observational prospective study recruiting consecutive patients that underwent zero-fluoroscopy LBBP pacemaker implantation. A 6-month follow-up visit was programmed for every patient. The main goal of our study was to assess the efficacy, feasibility, and safety of the procedure. RESULTS: From January 2021 to February 2022, we included 10 patients, 8 males. The average age was 63 ± 4 years. The procedure was successful in all patients. We observed a significant reduction in paced QRS width compared with basal QRS width (149 ± 31.9 vs. 116 ± 15.6 ms, p = .02). All device parameters remained stable at 6-month follow-up: no significant differences in mean impedance (700.5 ± 136.4 vs. 494 ± 72.7 Ohm, p = .09), capture threshold (0.67 ± 0.2 vs. 0.83 ± 0.2 V @ 0.4 ms, p = .27) or endocardial V-wave amplitude (10.6 ± 5.2 vs. 13.9 ± 6.3 mV, p = .19). No complications were reported in any case. CONCLUSION: Zero-fluoroscopy LBBP is feasible and safe, and it may be considered in cases where radiation exposure is contraindicated or especially undesirable. Future randomized clinical trials are needed for the widespread use of this new technique.
No takes yet. Share an insight, caveat, or question.
Ramos‐Maqueda et al. (2022) conducted an observational in Pacemaker implantation (n=10). Zero-fluoroscopy left bundle branch pacing (LBBP) was evaluated on Efficacy, feasibility, and safety of the procedure. Zero-fluoroscopy left bundle branch pacing was successful in 100% of patients, significantly reducing paced QRS width compared to basal width (116 vs 149 ms, p=0.02) with no complications at 6 months.