Key result
Pentaspline PFA during PVI unintentionally creates bidirectional mitral isthmus block due to suboptimal sheath alignment.
Case Report (n=1)
Unintentional mitral isthmus block can occur during PVI with a pentaspline PFA catheter due to suboptimal sheath-catheter alignment, highlighting the need for careful anatomical orientation.
Warrants caution for unintended mitral isthmus block during pentaspline PFA PVI; leaves open optimal alignment and clinical relevance.
This case demonstrates an unintentional bidirectional mitral isthmus block created during pulmonary vein isolation using a pentaspline pulsed-field ablation catheter. The finding underscores the importance of careful sheath–catheter alignment at the left inferior pulmonary vein to prevent unintended lesion extension toward the mitral annulus. Pulsed-field ablation (PFA) is a novel nonthermal technique for atrial fibrillation (AF) ablation that selectively targets myocardial cells while sparing surrounding noncardiac tissue. Recent clinical studies and real-world data have demonstrated the procedural efficacy and reproducibility of pulmonary vein isolation (PVI) using the pentaspline PFA catheter (FARAWAVE, Boston Scientific, Marlborough, MA, USA) [1, 2]. In Japan, the clinical use of PFA is currently limited to PVI; however, in other countries, previous studies have reported the creation of mitral isthmus block lines using the FARAPULSE system [2, 3]. This typically required additional applications delivered in a flower configuration at the mitral isthmus. Here, we describe a unique case in which bidirectional mitral isthmus block was unintentionally achieved with a minimal number of applications. A 60-year-old man with symptomatic paroxysmal AF underwent PFA using the FARAPULSE system (Boston Scientific) under general anesthesia (Figure 1). Following transseptal puncture using a 16.8-Fr PFA sheath (Faradrive, Boston Scientific) and a radiofrequency guidewire (Versacross, Boston Scientific), left atrial angiography was performed with a pigtail catheter during right ventricular burst pacing (Videos S1 and S2). PVI was performed under fluoroscopic guidance (without 3D mapping) with the minimum number of applications: eight per pulmonary vein (four in the flower and four in the basket configuration, for a total of 32 applications). Post-ablation mapping using the Ensite X system (Abbott, Chicago, IL, USA) during pacing from the coronary sinus (CS) ostium confirmed successful PVI and, unexpectedly, revealed a complete mitral isthmus block line (Figure 2A). This high-density map, constructed using an HD Grid catheter (Abbott) with more than 2000 acquisition points, was considered sufficiently reliable to demonstrate that a true conduction block had indeed occurred across the mitral isthmus. Subsequent pacing from a mapping catheter positioned in the left atrial appendage showed a proximal-to-distal activation sequence in the CS (Figure 2B). These findings confirmed bidirectional conduction block across the mitral isthmus. No major ST-segment changes were observed during the procedure, which was completed without any complications. During the 6-month follow-up, the patient remained free of AF and atrial tachycardia. Although previous studies have reported the creation of a mitral isthmus block line using the FARAPULSE system, this typically required additional applications in a flower configuration at the mitral isthmus [2, 3]. To the best of our knowledge, this is the first reported case in which a bidirectional mitral isthmus block was achieved with a minimum number of applications (32 applications). A retrospective review of fluoroscopic images (Figure 3) and schematic illustration (Figure 4) showed that (1) the left superior pulmonary vein applications did not appear to have contributed to the creation of the mitral isthmus block and (2) the PFA applications in the basket configuration at the left inferior pulmonary vein (LIPV) were delivered slightly more anteriorly and proximally than those in the flower configuration. The cause of the unintentional lesion formation is considered to be related to suboptimal alignment between the FARADRIVE sheath and the FARAWAVE catheter during the LIPV applications. In Figure 3, the tip of the FARADRIVE sheath is clearly oriented toward the anterior wall, suggesting that the catheter was inadvertently pulled anteriorly, thereby extending the application toward the lateral mitral isthmus. This misalignment would have been easier to recognize in the right anterior oblique view, whereas in the AP view it can remain subtle when using the basket configuration. Particularly during applications at the LIPV, it would have been preferable to reposition the CS catheter into the CS to serve as an anatomical landmark. Additionally, the naturally short length (18 mm in computed tomography, Figure 1B; 20 mm in 3D mapping, Figure 2A) of the mitral isthmus may have facilitated this unexpected outcome. The durability of the lesion formation in the LIPV and the unintentional mitral isthmus block line remains uncertain, as neither AF nor atrial tachycardia has recurred, and no repeat ablation procedure has been undertaken. This case highlights the importance of careful attention to catheter–sheath alignment and anatomical orientation during PFA applications at the LIPV to prevent unintentional mitral isthmus lesion formation—a phenomenon that cannot occur with PVI performed using radiofrequency ablation. We thank Dr. Jungo Kasai for editing a draft of this manuscript. This research was conducted according to the principles of the Declaration of Helsinki. The patient provided written informed consent for publication of the details of her case. The authors declare no conflicts of interest. The authors confirm that all the data supporting the findings of this research is available within the article. Video S1: Left atrial angiography performed with a pigtail catheter during right ventricular burst pacing (anteroposterior view). Video S2: Left atrial angiography performed with a pigtail catheter during right ventricular burst pacing (left anterior oblique view). Please note: The publisher is not responsible for the content or functionality of any supporting information supplied by the authors. Any queries (other than missing content) should be directed to the corresponding author for the article.
No takes yet. Share an insight, caveat, or question.
Kasai et al. (2025) conducted a case report in Symptomatic paroxysmal atrial fibrillation (n=1). Pulsed-field ablation using a pentaspline catheter (FARAPULSE system) was evaluated on Bidirectional mitral isthmus block. Pulmonary vein isolation using a pentaspline pulsed-field ablation catheter unintentionally created a bidirectional mitral isthmus block due to suboptimal sheath-catheter alignment.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: