Key result
Varipulse pulsed field ablation catheter achieves successful anterior mitral line block alongside PVI.
Why the study?
Despite growing experience with pulsed field ablation for pulmonary vein isolation, evidence was lacking regarding its use for linear ablation lesions.
Can the Varipulse PFA catheter be used to perform linear ablation lesions (anterior mitral line) in addition to pulmonary vein isolation in a patient with paroxysmal atrial fibrillation?
Case Report (n=1)
No
Can the Varipulse PFA catheter be used to perform linear ablation lesions (anterior mitral line) in addition to pulmonary vein isolation in a patient with paroxysmal atrial fibrillation?
The Varipulse PFA catheter can safely and effectively create linear atrial lesions in addition to pulmonary vein isolation, demonstrating its flexibility for complex left atrial substrate modification.
Background Pulsed field ablation (PFA) is a recently developed technology that causes tissue necrosis through electroporation and has been employed and validated for pulmonary vein isolation (PVI) in atrial fibrillation (AF) ablation. Only few devices have received a CE mark for this use and the VaripulseTM catheter (VaripulseTM, Biosense Webster, Irvine, USA), a variable loop PFA catheter, has been recently approved. Despite the growing experience with PVI, evidence is lacking about the use of PFA catheters and linear ablation lesions, which could potentially be performed with PFA catheters, reducing procedure related costs and time. Case summary A 73-year-old Caucasian woman with multiple cardiovascular risk factors presented with highly symptomatic paroxysmal AF and was scheduled to undergo PFA AF ablation using the new Varipulse catheter. The procedure was performed under deep sedation and the catheters were advanced through trans-septal puncture. 3D electroanatomic mapping of the left atrium showed significant anterior wall fibrosis. After PVI, substrate modification through an anterior mitral line (AML) approach was pursued. The AML was performed with additional PFA applications and successful bidirectional block was confirmed by pacing maneuvers without any procedural complications. The patient was discharged the following day, asymptomatic and in a state of well-being. Conclusion Individual LA substrate modification, including linear lesion sets, can be delivered in addition to PVI using the novel Varipulse PFA catheter in patients with AF.
No takes yet. Share an insight, caveat, or question.
Parlato et al. (2025) conducted a case report in Paroxysmal atrial fibrillation (n=1). Varipulse PFA catheter (pulsed field ablation) was evaluated on Acute procedural success (pulmonary vein isolation and bidirectional block of the anterior mitral line). Substrate modification with an anterior mitral line was successfully performed in addition to pulmonary vein isolation using the Varipulse pulsed field ablation catheter without complications.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: