Key result
Novel Re-PASTE technique safely enables right inferior PVI during pulsed field ablation.
Why the study?
Isolating the right inferior pulmonary vein during pulsed field ablation can be technically challenging due to anatomical limitations and restricted catheter maneuverability.
Case Report (n=1)
No
The Re-PASTE technique provides a reproducible and practical fluoroscopy-guided method to overcome anatomical challenges when isolating the right inferior pulmonary vein with a pentaspline pulsed field ablation catheter.
Answers are grounded in this paper and cite their sources.
May facilitate right inferior pulmonary vein isolation in challenging anatomies during pulsed field ablation; leaves open confirmation in prospective studies.
Pulsed field ablation is a nonthermal ablation modality with favorable safety and efficacy profiles for pulmonary vein isolation in patients with atrial fibrillation. However, isolating the right inferior pulmonary vein can be technically challenging because of anatomical limitations and restricted catheter maneuverability. We report the case of a 47-year-old man with symptomatic, drug-refractory paroxysmal atrial fibrillation who underwent successful pulsed field ablation under general anesthesia using the FARAPULSE system. The Retracting the PetAls (in the flower configuration) againST the sEptum (Re-PASTE) technique was used in combination with radiopaque markers on the pentaspline FARAWAVE catheter and FARADRIVE sheath to facilitate accurate right inferior pulmonary vein access using fluoroscopic guidance alone. The Re-PASTE technique benefits from the design features of the pentaspline PFA catheter and sheath, such as their widths and lengths and their radiopaque markers, thereby enabling clear identification of the right-left atrial boundary and stable catheter positioning without spline deformation. The technique is simple, reproducible, and effective in addressing right inferior pulmonary vein anatomical challenges. The Re-PASTE technique may enhance procedural safety and efficiency, particularly in cases with difficult right inferior pulmonary vein anatomy, offering a practical solution to a common technical limitation in pulsed field ablation procedures.
No takes yet. Share an insight, caveat, or question.
Kasai et al. (2025) conducted a case report in Paroxysmal atrial fibrillation (n=1). Re-PASTE technique with pulsed field ablation was evaluated on Successful right inferior pulmonary vein isolation and freedom from atrial fibrillation recurrence. The Re-PASTE technique safely and effectively enabled right inferior pulmonary vein isolation during pulsed field ablation, resulting in no atrial fibrillation recurrence over a 12-month follow-up.
Synapse has enriched 10 related papers. Consider them for comparative context: