Key result
High intensity focused ultrasound (HIFU) ablation failed to achieve sustained conduction block in rabbit atrial preparations due to acoustic radiation force and cavitation.
Why the study?
What are the mechanisms of failure for high intensity focused ultrasound (HIFU) ablation in cardiac tissue?
What are the mechanisms of failure for high intensity focused ultrasound (HIFU) ablation in cardiac tissue?
HIFU ablation failure in cardiac tissue is driven by acoustic radiation force, acoustic cavitation, and inconsistent thermal deposition leading to lesion discontinuity.
Answers are grounded in this paper and cite their sources.
HIFU should not yet enter clinical cardiac ablation trials; leaves open parameter optimization to mitigate acoustic tissue effects in models.
BACKGROUND: High intensity focused ultrasound (HIFU) has been introduced for treatment of cardiac arrhythmias because it offers the ability to create rapid tissue modification in confined volumes without directly contacting the myocardium. In spite of the benefits of HIFU, a number of limitations have been reported, which hindered its clinical adoption. METHODS AND RESULTS: In this study, we used a multimodal approach to evaluate thermal and nonthermal effects of HIFU in cardiac ablation. We designed a computer controlled system capable of simultaneous fluorescence mapping and HIFU ablation. Using this system, linear lesions were created in isolated rabbit atria (n=6), and point lesions were created in the ventricles of whole-heart (n=6) preparations by applying HIFU at clinical doses (4-16 W). Additionally, we evaluate the gap size in ablation lines necessary for conduction in atrial preparations (n=4). The voltage sensitive dye di-4-ANEPPS was used to assess functional damage produced by HIFU. Optical coherence tomography and general histology were used to evaluate lesion extent. Conduction block was achieved in 1 (17%) of 6 atrial preparations with a single ablation line. Following 10 minutes of rest, 0 (0%) of 6 atrial preparations demonstrated sustained conduction block from a single ablation line. Tissue displacement of 1 to 3 mm was observed during HIFU application due to acoustic radiation force along the lesion line. Additionally, excessive acoustic pressure and high temperature from HIFU generated cavitation, causing macroscopic tissue damage. A minimum gap size of 1.5 mm was found to conduct electric activity. CONCLUSIONS: This study identified 3 potential mechanisms responsible for the failure of HIFU ablation in cardiac tissues. Both acoustic radiation force and acoustic cavitation, in conjunction with inconsistent thermal deposition, can increase the risk of lesion discontinuity and result in gap sizes that promote ablation failure.
No takes yet. Share an insight, caveat, or question.
Laughner et al. (2012) studied Cardiac arrhythmias (n=16). High intensity focused ultrasound (HIFU) ablation was evaluated on Sustained conduction block. High intensity focused ultrasound (HIFU) ablation failed to achieve sustained conduction block in rabbit atrial preparations due to acoustic radiation force and cavitation.