Key result
Wrist-worn PPG strongly correlates with ECG for HR recovery despite lagging in ~97% of cases.
Why the study?
Heart rate accuracy from wrist-worn photoplethysmography is altered during exercise transitions, questioning the validity of estimated postexercise heart-rate-recovery parameters.
Does a wrist-worn photoplethysmography monitor accurately measure postexercise heart rate recovery compared to a 10-lead ECG?
Observational (n=30)
Randomized sequence
Does a wrist-worn photoplethysmography monitor accurately measure postexercise heart rate recovery compared to a 10-lead ECG?
Wrist-worn PPG monitors can assess heart rate recovery accurately if performed on a bicycle ergometer focusing on Δ60 and adjusting for the time lag in HR decrease.
PURPOSE: The accuracy of heart rate (HR) measured with a wrist-worn photoplethysmography (PPG) monitor is altered during rest-exercise and exercise-rest transitions, which questions the validity of postexercise HR-recovery (HRR) parameters estimated from this device. METHODS: Thirty participants (50% female) randomly performed two 13-minute sequences (3' rest, 5' submaximal-intensity exercise, and 5' passive recovery) on treadmill and bicycle ergometers. HR was measured concomitantly with a 10-lead electrocardiogram (ECG) and a wrist-worn PPG monitor (Polar Unite). HRR was assessed by calculating Δ60 (the difference between HR during exercise and HR 60 s after exercise cessation) and by fitting HRR data into a monoexponential model. RESULTS: By focusing on Δ60 and τ (the time constant of the monoexponential curve), levels of association (r) of the Unite versus the 10-lead ECG were high to very high (.73 < r < .93), and coefficients of variation were >20% (in absolute value), except for Δ60 in the bicycle ergometer condition (11.7%). In 97% of cases, the decrease in HR after exercise appeared later with the Unite. By adjusting the time window used for the analysis according to this time lag, coefficients of variation of Δ60 decreased below 10% in the bicycle ergometer condition. CONCLUSIONS: If a wrist-worn PPG monitor is used to assess HRR, we recommend performing the submaximal-intensity exercise on a bicycle ergometer and focusing on Δ60. Furthermore, to obtain a more accurate Δ60, the time lag between the end of the exercise and the effective decrease in HR should also be considered before the calculation.
No takes yet. Share an insight, caveat, or question.
Bretonneau et al. (2023) conducted an observational in Healthy participants (implied) (n=30). Wrist-worn PPG monitor (Polar Unite) vs. 10-lead electrocardiogram (ECG) was evaluated on Accuracy of heart-rate-recovery parameters (Δ60 and τ) (r = 0.73 to 0.93). A wrist-worn PPG monitor (Polar Unite) showed high association with 10-lead ECG for heart-rate recovery parameters (.73 < r < .93), though a time lag in HR decrease was observed in 97% of cases.