Key result
Audible steam pops complicate ~3% of RFCA for PVI procedures and feature abrupt temperature changes.
Why the study?
Steam pop during RFCA for PVI can cause cardiac perforation, but data on timely detection of silent or nonaudible steam pop events are limited.
Does continuous monitoring of impedance, temperature, and contact force allow for timely detection of steam pop events during radiofrequency catheter ablation in patients with atrial fibrillation?
Observational (n=516)
Does continuous monitoring of impedance, temperature, and contact force allow for timely detection of steam pop events during radiofrequency catheter ablation in patients with atrial fibrillation?
Simultaneous recording of spike potentials, abrupt impedance rise, and concomitant temperature or contact force changes may enable timely detection of potentially dangerous steam pop events during radiofrequency ablation.
Answers are grounded in this paper and cite their sources.
Monitoring temperature changes may aid steam pop detection during PVI; leaves open whether routine use improves safety in AF ablation.
Aim: Steam pop (SP) during radiofrequency catheter ablation (RFCA) for pulmonary vein isolation (PVI) may cause cardiac perforation, which may require drainage and emergent thoracotomy or even lead to death. Data investigating the timely detection of the occurrence of "silent" or "nonaudible" SP events are limited. Methods and Results: A total of 516 consecutive atrial fibrillation (AF) patients who underwent index PVI were included in this retrospective observational study. The duration, power, impedance, temperature, and contact force (CF) of RFCA were continually monitored and recorded throughout the procedure. A total of 15 (2.9%) audible SP events occurred in 14 patients; 2 of the patients developed pericardial tamponade, 1 patient underwent drainage, and 1 patient underwent emergent thoracotomy. The time from RFCA initiation to the occurrence of audible SP was 19.4 ± 6.9 s. Abrupt temperature change occurred in 13 (86.7%) of the 15 SP events, of which 8 (53.3%) exhibited an abrupt temperature rise of 2.3 ± 1.0°C, 5 (33.3%) exhibited an abrupt temperature drop of 2.3 ± 1.3°C, and 2 (13.3%) exhibited no discernible temperature change. Conclusions: In conclusion, simultaneously recorded spike potentials and abrupt impedance rise with concomitant temperature and/or CF change could be a feasible method for the timely detection of the occurrence of audible, "silent," or "nonaudible" SP events, particularly in regions where the risk of perforation may be of concern.
No takes yet. Share an insight, caveat, or question.
Di et al. (2023) conducted an observational in Atrial fibrillation (n=516). Radiofrequency catheter ablation (RFCA) for pulmonary vein isolation (PVI) was evaluated on Audible steam pop (SP) events. Audible steam pop events occurred in 2.9% of patients undergoing RFCA for PVI, with 86.7% of events exhibiting abrupt temperature changes, suggesting monitoring may aid timely detection.