Answers are grounded in this paper and cite their sources.
LOD and LOQ are two important performance characteristics in method validation. This work compares three methods based on the International Conference on Harmonization and EURACHEM guidelines, namely, signal-to-noise, blank determination, and linear regression, to estimate the LOD and LOQ for volatile organic compounds (VOCs) by experimental methodology using GC. Five VOCs, toluene, ethylbenzene, isopropylbenzene, n-propylbenzene, and styrene, were chosen for the experimental study. The results indicated that the estimated LODs and LOQs were not equivalent and could vary by a factor of 5 to 6 for the different methods. It is, therefore, essential to have a clearly described procedure for estimating the LOD and LOQ during method validation to allow interlaboratory comparisons.
No takes yet. Share an insight, caveat, or question.
Sanagi et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: