Key result
The proposed method using ECG, PPG, and intermittent measurements estimated systolic blood pressure every 30 seconds with a mean difference of 0.08 mmHg and standard deviation of difference of 7.97 mmHg.
Why the study?
Intermittent non-invasive blood pressure measurement during surgery carries a risk of missing clinically relevant blood pressure changes between measurements.
Does a dynamic feature selection method using ECG, PPG, and intermittent BP measurements accurately estimate continuous systolic blood pressure in patients undergoing surgery?
Observational (n=29)
No
Does a dynamic feature selection method using ECG, PPG, and intermittent BP measurements accurately estimate continuous systolic blood pressure in patients undergoing surgery?
Absolute Event Rate: 7.97% vs 13.12%
p-value: p=<0.001
A dynamic feature selection model using ECG, PPG, and intermittent BP measurements can accurately estimate continuous systolic blood pressure during surgery, meeting AAMI standards.
May support continuous non-invasive SBP monitoring in surgery; leaves open prospective validation before practice change.
In a significant portion of surgeries, blood pressure (BP) is often measured non-invasively in an intermittent manner. This practice has a risk of missing clinically relevant BP changes between two adjacent intermittent BP measurements. This study proposes a method to non-invasively estimate systolic blood pressure (SBP) with high accuracy in patients undergoing surgery. Continuous arterial BP, electrocardiography (ECG), and photoplethysmography (PPG) signals were acquired from 29 patients undergoing surgery. After extracting 9 features from the PPG and ECG signals, we dynamically selected features upon each intermittent measurement (every 10 min) of SBP based on feature robustness and the principle of correlation-based feature selection. Finally, multiple linear regression models were built to combine these features to estimate SBP every 30 s. Compared to the reference SBP, the proposed method achieved a mean of difference at 0.08 mmHg, a standard deviation of difference at 7.97 mmHg, and a correlation coefficient at 0.89 (p < 0.001). This study demonstrates the feasibility of non-invasively estimating SBP every 30 s with high accuracy during surgery by using ECG, PPG, and intermittent SBP measurements every 10 min, which meets the standard of the Association for the Advancement of Medical Instrumentation. The proposed method has the potential to enhance BP monitoring in the operating room, improving patient outcomes and experiences.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2022) conducted an observational in Patients undergoing major surgery (n=29). Dynamic feature selection using ECG, PPG, and intermittent SBP measurements vs. Zero-order hold method was evaluated on Systolic blood pressure estimation accuracy (pooled RMSE in mmHg) (p=<0.001). The proposed method using ECG, PPG, and intermittent measurements estimated systolic blood pressure every 30 seconds with a mean difference of 0.08 mmHg and standard deviation of difference of 7.97 mmHg.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: