Key result
Rampart M1128 system is noninferior to standard lead aprons for STEMI D2B time.
Why the study?
Enhanced radiation protection systems reduce occupational hazards and orthopedic injuries from lead aprons, but their impact on door-to-balloon time in STEMI has not been evaluated.
Does the Rampart M1128 enhanced radiation protection system prolong door-to-balloon time compared to standard lead aprons in STEMI patients undergoing percutaneous coronary intervention?
Cohort (n=174)
No
Does the Rampart M1128 enhanced radiation protection system prolong door-to-balloon time compared to standard lead aprons in STEMI patients undergoing percutaneous coronary intervention?
The use of a novel radiation protection system (Rampart M1128) does not delay door-to-balloon times during STEMI interventions, supporting its adoption to improve operator safety without compromising patient care.
Noninferior D2B times with Rampart in STEMI PCI cohort; leaves open randomized confirmation before adoption.
Background: Prolonged radiation exposure remains a critical occupational hazard in the catheterization lab, typically mitigated by lead aprons, which contribute to orthopedic injury. Recent innovations like the Rampart M1128 enhanced radiation protection system (Rampart IC) aim to reduce these risks, but their impact on door-to-balloon (D2B) time in ST-elevation myocardial infarction (STEMI) has not been evaluated. Methods: We conducted a single-center retrospective cohort study comparing D2B times among 174 STEMI patients undergoing percutaneous coronary intervention: 87 treated using lead aprons and 87 treated utilizing the Rampart M1128 system. Patients were randomly selected from a pool of 298. Analyses were adjusted for potential confounders, including bleeding, use of vasopressors, intubation, mechanical support, electrical instability, access type, and prior revascularization. Results: = .21), well within the prespecified 5-minute noninferiority margin. No statistically significant differences were observed across expanded models or one-at-a-time covariate models. The presence of shock or respiratory failure did not prolong D2B times in the Rampart group. Conclusions: Implementation of an enhanced radiation protection system was noninferior to standard lead apron use regarding D2B time in STEMI patients. This suggests such systems can enhance operator safety without compromising the quality of patient care. Broader adoption may be justified to reduce occupational harm while maintaining high-quality cardiovascular outcomes.
No takes yet. Share an insight, caveat, or question.
Hallak et al. (2025) conducted a cohort in ST-elevation myocardial infarction (STEMI) (n=174). Rampart M1128 enhanced radiation protection system vs. Standard lead aprons was evaluated on Door-to-balloon (D2B) time. The Rampart M1128 enhanced radiation protection system was noninferior to standard lead aprons for door-to-balloon time in STEMI patients, falling within the prespecified 5-minute margin.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: