Comparative laboratory study reveals that heat-treated murine immunoglobulin and Fc fragment removal reduce interference in serum assays, indicating practical solutions for diagnostic accuracy.
BACKGROUND: The primary aim of the study was to reduce interference in an in-house two-site, two-step immunometric assay. METHODS: In the running laboratory routine, 11 261 samples were tested with a carcinoembryonic antigen (CEA) assay with bovine immunoglobulin but no murine immunoglobulins in the buffer, in parallel to our routine CEA assay, using 15 mg/L heat-treated nonspecific murine immunoglobulin (MAK33) in the buffer and with the Fc fragments removed from the capture antibody. RESULTS: The frequency of interference was estimated to be 4.0% (95% confidence interval, 3.3-4.7%). The addition of 15 mg/L native MAK33 had little effect (frequency, 3.9%; 95% confidence interval, 3.2-4.6%), whereas adding 15 mg/L heat-treated MAK33 reduced interference to 0.86% (0.61-1.12%), and adding 50 mg/L reduced it further to 0.06% (0-0.13%). Removing the Fc fragments by itself reduced interference to 0.10% (0.02-0.19%). There were no statistically significant differences for age (P <0.23) or gender (P <0.40) between patients with interference (n = 210) and a randomly selected interference-negative control group (n = 186). Interference was not constant in patients: 15 of 25 individuals positive for interference and with four or more samples screened for interference had an interference-negative sample either before or after the peak of interference. CONCLUSIONS: In a two-site, two-step immunometric assay using mouse monoclonal antibodies, use of heat-treated nonspecific murine immunoglobulin in the buffer or removal of the Fc fragment from the capture antibody could improve performance.
No takes yet. Share an insight, caveat, or question.
Bjerner et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: