BACKGROUND: Heterophilic antibody (HA) interference remains a persistent challenge in immunoassay diagnostics. Human anti-mouse antibodies (HAMA) can cause spuriously elevated biomarker results, potentially leading to misdiagnosis and unnecessary clinical interventions. Brain natriuretic peptide (BNP) is critical for heart failure diagnosis, but false-positive results due to HAMA may trigger inappropriate clinical management. CASE PRESENTATION: A 53 year-old male with type 2 diabetes presented with markedly elevated BNP (4394-4419 pg/mL; reference < 100 pg/mL) by chemiluminescence immunoassay, despite no heart failure symptoms, normal ejection fraction (62%), and normal NT-proBNP (55.8 pg/mL). This discordance prompted investigation for analytical interference. METHODS: Serial dilution (1:4, 1:8); specific blocking with mouse IgG, goat IgG, polymeric antibodies, and RF blocker; dose-response titration of mouse IgG (200-600 μg/mL); and alternative platform comparison. Mouse IgG induced a modest but specific 15.7% reduction in apparent BNP. Even 600 μg/mL mouse IgG achieved only partial blocking (residual BNP 2240 pg/mL), indicating high-titer HAMA exceeding standard blocking capacity. CONCLUSIONS: This case demonstrates extreme HAMA interference causing false-positive BNP > 4000 pg/mL. Standard blocking reagents may be insufficient for high-titer samples. When BNP is discordant with clinical status-especially with normal NT-proBNP-laboratories must use systematic verification to prevent misdiagnosis.