Key result
Adding purified mouse IgG normalizes falsely elevated BNP, confirming human anti-mouse antibody interference.
Why the study?
BNP immunoassays are vulnerable to heterophilic antibody interference that can cause false-positive elevations and clinical mismanagement, accounting for unexplained biomarker elevations.
Does a systematic laboratory investigation identify and correct falsely elevated BNP due to HAMA interference?
Case Report (n=1)
Does a systematic laboratory investigation identify and correct falsely elevated BNP due to HAMA interference?
A systematic laboratory workflow can identify and correct falsely elevated BNP levels caused by human anti-mouse antibody interference, preventing unnecessary clinical interventions.
No takes yet. Share an insight, caveat, or question.
Suggests evaluating for HAMA in unexplained BNP elevations; confirms interference mechanism but leaves prevalence and impact open.
Li et al. (2026) conducted a case report in Falsely elevated B-type natriuretic peptide (n=1). Purified mouse IgG was evaluated on BNP level normalization. Addition of purified mouse IgG normalized the falsely elevated BNP level (2735.7 ± 54.3 pg/mL) in a dose-dependent manner, confirming human anti-mouse antibody interference.