Key result
Pulsed electric field delivery to porcine airways induces no sustained arrhythmias, even during the T-wave.
Why the study?
Differences between pulsed electric field systems make cross-technology cardiac safety assessment challenging, though evidence suggests shorter biphasic pulses may obviate the need for cardiac cycle synchronization.
Does monopolar, biphasic, microsecond-scale PEF delivery to the airways induce cardiac arrhythmias in a porcine model?
Does monopolar, biphasic, microsecond-scale PEF delivery to the airways induce cardiac arrhythmias in a porcine model?
Short-duration, biphasic, monopolar pulsed electric field therapy delivered to the airways does not induce clinically meaningful cardiac arrhythmias in a porcine model, suggesting cardiac synchronization may not be required.
Answers are grounded in this paper and cite their sources.
May obviate cardiac synchronization for airway PEF; hypothesis-generating pending human validation.
Pulsed electric field (PEF) technologies treat many types of tissue. Many systems mandate synchronization to the cardiac cycle to avoid the induction of cardiac arrhythmias. Significant differences between PEF systems make the assessment of cardiac safety from one technology to another challenging. A growing body of evidence suggests that shorter duration biphasic pulses obviate the need for cardiac synchronization, even when delivered in a monopolar fashion. This study theoretically evaluates the risk profile of different PEF parameters. It then tests a monopolar, biphasic, microsecond-scale PEF technology for arrhythmogenic potential. PEF applications of increasing likelihood to induce an arrhythmia were delivered. The energy was delivered throughout the cardiac cycle, including both single and multiple packets, and then with concentrated delivery on the t-wave. There were no sustained changes to the electrocardiogram waveform or to the cardiac rhythm, despite delivering energy during the most vulnerable phase of the cardiac cycle, and delivery of multiple packets of PEF energy across the cardiac cycle. Only isolated premature-atrial contractions (PAC) were observed. This study provides evidence that certain varieties of biphasic, monopolar PEF delivery do not require synchronized energy delivery to prevent harmful arrhythmias.
No takes yet. Share an insight, caveat, or question.
Meininger et al. (2023) studied Healthy swine (n=4). Monopolar, biphasic pulsed electric fields (PEF) was evaluated on Sustained changes to cardiac rhythm or ECG waveform (e.g., ventricular fibrillation, tachycardia). Asynchronous delivery of monopolar, biphasic pulsed electric fields to porcine airways did not induce any sustained cardiac arrhythmias across 3,125 delivered packets, even during the vulnerable T-wave.
Synapse has enriched 10 related papers. Consider them for comparative context: