Key result
Transesophageal echocardiography using the double-envelope technique accurately calculated prosthetic aortic valve area, correlating well with manufacturer data.
Why the study?
Does the double-envelope technique using TEE accurately calculate prosthetic aortic valve area compared to manufacturer data in patients immediately after aortic valve replacement?
Observational (n=45)
Does the double-envelope technique using TEE accurately calculate prosthetic aortic valve area compared to manufacturer data in patients immediately after aortic valve replacement?
The double-envelope technique using continuous wave Doppler during TEE provides an accurate, simultaneous measurement of subvalvular and valvular velocities to calculate prosthetic aortic valve area, simplifying the continuity equation.
Answers are grounded in this paper and cite their sources.
May support simplified prosthetic AVA assessment by TEE post-AVR; hypothesis-generating and requires prospective validation before practice change.
UNLABELLED: The conventional continuity equation uses nonsimultaneous measurements of blood flow velocities through the left ventricular outflow tract and across the aortic valve to calculate aortic valve area (AVA). We have noted that both velocities can be simultaneously obtained from continuous wave (CW) Doppler analysis (double-envelope [DE]). We hypothesize that prosthetic AVA can be calculated by using the DE technique, during transesophageal echocardiography (TEE). Prosthetic AVA was calculated in 41 of 45 patients immediately after aortic valve replacement by using the DE/AVA technique. Left ventricular outflow tract diameter was obtained from an esophageal view, while subvalvular (V(1)) and valvular (V(2)) peak velocities were simultaneously obtained from transgastric views by using CW Doppler. Prosthetic AVA and V(1)/V(2) ratio (Doppler velocity index) were calculated. V(1) was also measured by using pulse wave Doppler, as is conventionally done. Twenty-three Carbomedic (CM) and 18 Carpentier-Edwards (CE) AVA were evaluated. DE/AVAs for CM and CE valves correlated and agreed with that reported by the manufacturer (CM r(2) = 0.91, mean bias -0.25 cm(2) [SD 0.18]; CE r(2) = 0.73, mean bias -0.02 cm(2) [SD 0.27]). Calculated Doppler velocity index values agree with available data (mean bias 0.03 [SD 0.05]). The V(1) obtained by using the DE method was nearly identical to the V(1) obtained by using pulse wave (r(2) = 0.95, mean bias 0.02 m/s [SD 0.04 m/s]). TEE assessment of prosthetic AVA using the DE technique agrees with data reported by the manufacturer. Obtaining subvalvular and valvular velocities from the same CW Doppler trace may simplify the continuity equation and help avoid errors caused by beat-to-beat changes in blood flow. Quantitative prosthetic aortic valve assessment can be performed, on-line, with TEE by using the DE technique. IMPLICATIONS: Quantitative assessment of prosthetic aortic valve area can be performed on-line by using transesophageal echocardiography using the double envelope technique.
No takes yet. Share an insight, caveat, or question.
Maslow et al. (2000) conducted an observational in Prosthetic aortic valves (n=45). Double-envelope (DE) technique using transesophageal echocardiography vs. Manufacturer reported data and conventional pulse wave Doppler was evaluated on Correlation and agreement of prosthetic aortic valve area (AVA) with manufacturer reported data (CM r2 = 0.91; CE r2 = 0.73). Transesophageal echocardiography using the double-envelope technique accurately calculated prosthetic aortic valve area, correlating well with manufacturer data.