Key result
Transcatheter aortic valve prostheses demonstrated a mean effective orifice area of 1.82±0.43 cm2 and a mean gradient of 9.3±4.5 mmHg, indicating mild obstruction compared to physiological valves.
Observational (n=146)
This study establishes normal reference ranges for Doppler hemodynamics and effective orifice areas for Edwards SAPIEN and CoreValve transcatheter aortic valves, demonstrating mild obstruction but favorable profiles compared to surgical bioprostheses.
Supports TAVR echocardiographic assessment with new reference ranges; leaves open prospective outcome validation before routine adoption.
AIMS: Transcatheter aortic valve implantation (TAVI) is a new therapy for severe aortic stenosis in high-risk patients. So far, no reference values for the echocardiographic assessment of this new class of heart valves have been established. The aim of our study was to determine Doppler haemodynamics and the effective orifice area (EOA). METHODS AND RESULTS: We retrospectively analysed the earliest transthoracic echocardiographic examinations of 146 stable patients after successful TAVI (median 8±20 days). Doppler examinations were analysed for peak instantaneous velocity, peak, and the mean systolic gradient. EOA was determined using the continuity equation. Patients with severe paravalvular aortic or mitral valve regurgitation were excluded. The overall peak instantaneous velocity (n=146) was 2.0±0.4 m/s with a peak systolic gradient of 17.1±7.4 mmHg and a mean gradient of 9.3±4.5 mmHg. The mean EOA was 1.82±0.43 cm2 with an indexed EOA of 1.0±0.27 cm2/m2. In general, all prostheses showed similar values-with the exception of the Edwards Sapien 23 mm which was associated with higher velocities and peak pressure gradients. CONCLUSION: Our study establishes the normal range for Doppler haemodynamics of four transcatheter aortic valve prostheses. Compared with previously published data of surgically implanted bioprostheses percutaneous valves tend to have similar EOA values but lower mean peak velocities and pressure gradients. In comparison with physiological haemodynamics; however, this new class of heart valves is still associated with a mild obstruction.
No takes yet. Share an insight, caveat, or question.
Spethmann et al. (2012) conducted an observational in severe aortic stenosis (n=146). Transcatheter aortic valve implantation (Edwards SAPIEN and CoreValve) was evaluated on Doppler haemodynamics and effective orifice area (EOA). Transcatheter aortic valve prostheses demonstrated a mean effective orifice area of 1.82±0.43 cm2 and a mean gradient of 9.3±4.5 mmHg, indicating mild obstruction compared to physiological valves.