Key result
Aortic valve design alters hemodynamics, with Hancock II and Evolut producing the highest shear stress.
Why the study?
The study was conducted to characterize the flow dynamics of surgical and transcatheter aortic valves in vitro, assessing differences in flow turbulence, blood damage likelihood, and hemodynamic performance across valve types.
Answers are grounded in this paper and cite their sources.
In vitro flow analysis demonstrates that aortic valve prosthesis design and size significantly influence downstream turbulence and shear stresses, which are associated with structural valve deterioration and blood damage.
Objective: To perform an in vitro characterization of surgical aortic valves (SAVs) and transcatheter aortic valves (TAVs) to highlight the development of the flow dynamics depending on the type of valve implanted and assess the basic differences in the light of flow turbulence and its effect on blood damage likelihood and hemodynamic parameters that shed light on valve performance. Methods: A Starr-Edwards ball and cage valve of internal diameter 22 mm, a 23-mm Medtronic Hancock II SAV, a 23-mm St Jude Trifecta SAV, a 23-mm St Jude SJM (mechanical valve) SAV, a 26-mm Medtronic Evolut TAV, and a 26-mm Edwards SAPIEN 3 TAV were assessed in a pulse duplicator under physiological conditions. Particle image velocimetry was performed for each valve. Pressure gradient and effective orifice area (EOA) along with velocity flow field, Reynolds shear stress (RSS), and viscous shear stress (VSS) were calculated. The flow characteristics and behavior downstream of the valves differed depending on the valve type, design, and size. The greater the RSS and VSS the more turbulent the downstream flow. Hancock II displays the greatest range of RSS and VSS magnitudes compared with the same-size Trifecta and SJM. The Evolut displays the greatest range of RSS and VSS compared with the SAPIEN 3. Conclusions: The results of this study shed light on numerous advancements in the design of aortic valve replacement prosthesis and the subsequent hemodynamic variations. Future surgical and transcatheter valve designs should aim at not only concentrating on hemodynamic parameters but also at optimizing downstream flow properties.
No takes yet. Share an insight, caveat, or question.
Hatoum et al. (2022) studied In vitro assessment of aortic valve prostheses (n=6). Aortic valve prostheses (SAVs and TAVs) was evaluated on Mean transvalvular pressure gradient and effective orifice area (p=<.0001). Flow characteristics and downstream turbulence differed significantly by valve type and design, with Hancock II and Evolut displaying the greatest ranges of Reynolds and viscous shear stress magnitudes.