Why the study?
What are the size-specific effective orifice areas and hemodynamic performances of the Edwards Sapien 3 prosthesis up to 1 year post-TAVI?
What are the size-specific effective orifice areas and hemodynamic performances of the Edwards Sapien 3 prosthesis up to 1 year post-TAVI?
The Edwards Sapien 3 prosthesis demonstrates size-dependent effective orifice areas but comparable and stable transvalvular gradients up to 1 year post-TAVI.
May inform Sapien 3 sizing by expected EOA; extends short-term hemodynamic data but leaves durability open.
Aims: The Edwards Sapien 3 heart valve prosthesis (S3) is commonly used for transcatheter aortic valve implantation (TAVI) and is available in three sizes. To date no data has been published on the effective orifice area (EOA) and the hemodynamic performance of the three different S3 sizes. The aim of this study was to measure the size-specific EOA and hemodynamic performance of the S3 in short-term and 1-year follow-up. Methods and results: One hundred and thirteen consecutive patients treated by TAVI with a S3 prosthesis at the Heart Clinic Zurich between May 2014 and July 2015 were included. Clinical data were extracted from the Swiss TAVI registry. The EOA was calculated using Doppler echocardiography (peri-interventionally and at discharge) and by 3D-biplane transoesophageal echocardiography (peri-interventionally). Mean transvalvular gradients (dPmean) were additionally calculated with Doppler echocardiography at 30 days and 1 year. Results were analysed separately for the 23 mm (n = 42; 37%), 26 mm (n = 46; 41%), and 29 mm (n = 25; 22%) prostheses. At discharge, the EOAs were 1.6 ± 0.2 cm2 (23 mm S3), 2.0 ± 0.2 cm2 (26 mm S3), and 2.7 ± 0.2 cm2 (29 mm S3), p < 0.001. The dPmeans at discharge were 10.9 ± 6.0 mmHg (23 mm S3), 10.4 ± 3.5 mmHg (26 mm S3), and 8.9 ± 2.8 mmHg (29 mm S3), p = 0.235, and did not significantly change over time within any of the S3 sizes. Conclusions: Post-TAVI, the EOAs of the three different S3 prosthesis sizes differ significantly, the transvalvular gradients, however, are comparable. Mean transvalvular gradients remain stable over time and document good prosthesis function after 1 year.
No takes yet. Share an insight, caveat, or question.
Külling et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: