Key result
Directed Graph Mapping outperforms Phase Mapping in an in-silico model, correctly identifying ~15% more true rotors.
Why the study?
Phase mapping identifies true AF sources but also false sources from far-field effects and interpolation errors, prompting evaluation of directed graph mapping.
Does Directed Graph Mapping (DGM) improve the discrimination of true and false rotors compared to Phase Mapping (PM) in an in-silico model of atrial fibrillation?
Does Directed Graph Mapping (DGM) improve the discrimination of true and false rotors compared to Phase Mapping (PM) in an in-silico model of atrial fibrillation?
Directed Graph Mapping outperforms Phase Mapping in discriminating true from false rotors in a computer model of atrial fibrillation by adjusting conduction velocity parameters.
Supports further DGM development for AF rotor mapping; leaves open clinical superiority to phase mapping pending in vivo trials.
Atrial fibrillation (AF) is the most frequently encountered arrhythmia in clinical practise. One of the major problems in the management of AF is the difficulty in identifying the arrhythmia sources from clinical recordings. That difficulty occurs because it is currently impossible to verify algorithms which determine these sources in clinical data, as high resolution true excitation patterns cannot be recorded in patients. Therefore, alternative approaches, like computer modelling are of great interest. In a recent published study such an approach was applied for the verification of one of the most commonly used algorithms, phase mapping (PM). A meandering rotor was simulated in the right atrium and a basket catheter was placed at 3 different locations: at the Superior Vena Cava (SVC), the Crista Terminalis (CT) and at the Coronary Sinus (CS). It was shown that although PM can identify the true source, it also finds several false sources due to the far-field effects and interpolation errors in all three positions. In addition, the detection efficiency strongly depended on the basket location. Recently, a novel tool was developed to analyse any arrhythmia called Directed Graph Mapping (DGM). DGM is based on network theory and creates a directed graph of the excitation pattern, from which the location and the source of the arrhythmia can be detected. Therefore, the objective of the current study was to compare the efficiency of DGM with PM on the basket dataset of this meandering rotor. The DGM-tool was applied for a wide variety of conduction velocities (minimal and maximal), which are input parameters of DGM. Overall we found that DGM was able to distinguish between the true rotor and false rotors for both the SVC and CT basket positions. For example, for the SVC position with a CVmin=0.01cmms, DGM detected the true core with a prevalence of 82% versus 94% for PM. Three false rotors where detected for 39.16% (DGM) versus 100% (PM); 22.64% (DGM) versus 100% (PM); and 0% (DGM) versus 57% (PM). Increasing CVmin to 0.02cmms had a stronger effect on the false rotors than on the true rotor. This led to a detection rate of 56.6% for the true rotor, while all the other false rotors disappeared. A similar trend was observed for the CT position. For the CS position, DGM already had a low performance for the true rotor for CVmin=0.01cmms (14.7%). For CVmin=0.02cmms the false and the true rotors could therefore not be distinguished. We can conclude that DGM can overcome some of the limitations of PM by varying one of its input parameters (CVmin). The true rotor is less dependent on this parameter than the false rotors, which disappear at a CVmin=0.02cmms. In order to increase to detection rate of the true rotor, one can decrease CVmin and discard the new rotors which also appear at lower values of CVmin.
No takes yet. Share an insight, caveat, or question.
Nieuwenhuyse et al. (2021) studied Atrial fibrillation. Directed Graph Mapping (DGM) vs. Phase Mapping (PM) was evaluated on Detection of true core at the Superior Vena Cava (SVC) position with CVmin=0.01cm/ms. Directed Graph Mapping distinguished true from false rotors better than Phase Mapping in an in-silico model, detecting the true core at the SVC position in 82% vs 94% of cases.