Key result
The Masimo Radical-7 pulse co-oximeter was less accurate than the HemoCue system, with 40% of readings within 1 g/dL of laboratory values compared to 90% for HemoCue.
Why the study?
Does point-of-care haemoglobin measurement using Radical-7 pulse co-oximeter or HemoCue accurately reflect laboratory haemoglobin measurements in women undergoing caesarean section delivery?
Observational (n=137)
Does point-of-care haemoglobin measurement using Radical-7 pulse co-oximeter or HemoCue accurately reflect laboratory haemoglobin measurements in women undergoing caesarean section delivery?
Answers are grounded in this paper and cite their sources.
The HemoCue system is significantly more accurate than the Radical-7 pulse co-oximeter for point-of-care hemoglobin measurement in obstetric patients when compared to laboratory values.
We prospectively compared two point-of-care haemoglobin concentration measuring devices with laboratory measurements to determine their accuracy in women undergoing caesarean section delivery. The two devices were the Masimo Rainbow SET(®) Radical -7™ pulse co-oximeter and the HemoCue(®) HB 201+, which is a cuvette-type system that uses photometry. Co-oximeter readings and HemoCue measurements were taken before and after surgery, and compared with laboratory measurements of haemoglobin concentration taken at the same time. We analysed data from 137 patients using Bland-Altman plots. Limits of agreement for co-oximeter readings were -2.19 to 3.41 g.dl(-1) and for the HemoCue were -1.52 to 1.79 g.dl(-1) . The bias (mean difference) for the co-oximeter was -0.61 g.dl(-1) (95% CI 0.36 to -0.86) and for the HemoCue was 0.13 g.dl(-1) (95% CI -0.015 to 0.28). [corrected] Overall, 110/274 (40%) co-oximeter readings were within 1 g.dl(-1) of laboratory values compared with 247/274 (90%) HemoCue measurements (p < 0.001 for difference). The co-oximeter gave lower readings and was less accurate than the HemoCue system when compared with laboratory measurements.
No takes yet. Share an insight, caveat, or question.
Skelton et al. (2012) conducted an observational in Women undergoing caesarean section delivery (n=137). Masimo Rainbow SET Radical-7 pulse co-oximeter vs. HemoCue HB 201+ and laboratory measurements was evaluated on Readings within 1 g.dl(-1) of laboratory values (p=< 0.001). The Masimo Radical-7 pulse co-oximeter was less accurate than the HemoCue system, with 40% of readings within 1 g/dL of laboratory values compared to 90% for HemoCue.