September 1, 2014Anaesthesia and Intensive CareOpen Access

Comparison of Massimo Pronto-7 and Hemocue Hb 201+ with Laboratory Haemoglobin Estimation: A Clinical Study

View Full Paper
Ask AI
Bookmark
Share

Key result

Point-of-care haemoglobin estimation using Pronto-7 and HemoCue showed fixed biases of +1.18 g/dl and -0.01 g/dl compared to laboratory analysis, with wide 95% prediction limits for both devices.

Why the study?

Do point-of-care devices (Pronto-7 and HemoCue) accurately estimate hemoglobin compared to an automated laboratory analyzer in pregnant women?

Population

141 pregnant women undergoing routine mid-trimester haemoglobin assessment

Comparison

Point-of-care hemoglobin estimation using… vs Automated laboratory analyser (Sysmex XE5000)

Design

Cross-sectional

Authors

RHRichard HiscockMercy Hospital for WomenSSS. W. SimmonsTulane UniversityBCBendix CarstensenSteno Diabetes Centers

Discussion

Loading...

Member takes

Implication

Wide prediction limits advise caution with Pronto-7 and HemoCue hemoglobin estimates in pregnancy; leaves open need for prospective device validation before clinical adoption.

Study Design

Type

Observational (n=141)

Structured PICO

Do point-of-care devices (Pronto-7 and HemoCue) accurately estimate hemoglobin compared to an automated laboratory analyzer in pregnant women?

P
Population
141 pregnant women undergoing routine mid-trimester haemoglobin assessment.
E
Exposure
Point-of-care hemoglobin estimation using Pronto-7 Pulse CO-Oximetry and HemoCue Hb 201+
C
Comparator
Automated laboratory analyser (Sysmex XE5000)
O
Outcome
Agreement in haemoglobin estimation (repeatability, between person within method bias and precision)surrogate

Main Result

Mean Difference: 1.18

Both Pronto-7 and HemoCue point-of-care devices are imprecise with wide prediction limits for hemoglobin estimation compared to laboratory analysis.

Cite This Study

Hiscock et al. (2014) conducted an observational in Haemoglobin assessment in pregnancy (n=141). Pronto-7 and HemoCue point-of-care devices vs. Automated laboratory analyser was evaluated on Agreement in haemoglobin estimation (fixed bias) (MD +1.18 g/dl (Pronto-7) and -0.01 g/dl (HemoCue)). Point-of-care haemoglobin estimation using Pronto-7 and HemoCue showed fixed biases of +1.18 g/dl and -0.01 g/dl compared to laboratory analysis, with wide 95% prediction limits for both devices.

synapsesocial.com/papers/6a9620dfbfacb0dfa03590eahttps://doi.org/10.1177/0310057x1404200510
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Evaluation of point‐of‐care haemoglobin measuring devices: a comparison of Radical‐7™ pulse co‐oximetry, HemoCue ® and laboratory haemoglobin measurements in obstetric patients*2012 · 67 citations
  2. 2The assessment of methods of measurement1990 · 158 citations