Key result
Updated standards raise the MRI magnet mode threshold for active implantable devices to 0.9 mT.
Why the study?
The review was conducted to explore the history of the 5 gauss (0.5 mT) safety threshold, describe the background behind updating this threshold to 9 gauss (0.9 mT) in IEC 60601-2-33 edition 4.0, and discuss the practical implications.
The review outlines the rationale and implications of updating the MRI safety threshold for active implants from 0.5 mT to 0.9 mT to prevent inadvertent magnet mode activation.
Answers are grounded in this paper and cite their sources.
May broaden MRI access for patients with active implants; leaves open prospective validation of the revised threshold.
The "5 gauss line" is a phrase that is likely to be familiar to everyone working with MRI, but what is its significance, how was it defined, and what changes are currently in progress? This review explores the history of 5 gauss (0.5 mT) as a threshold for protecting against inadvertently putting cardiac pacemakers, implantable cardioverter defibrillators, and other active implantable medical devices into a "magnet mode." Additionally, it describes the background to the recent change of this threshold to 9 gauss (0.9 mT) in the International Standard IEC 60601-2-33 edition 4.0 that defines basic safety requirements for MRI. Practical implications of this change and some ongoing and emerging issues are also discussed.
No takes yet. Share an insight, caveat, or question.
Steckner et al. (2024) conducted a review in Active implantable medical devices (pacemakers, ICDs) in MRI. Threshold change from 0.5 mT to 0.9 mT was evaluated. The threshold for protecting active implantable medical devices from inadvertent magnet mode activation during MRI was recently changed from 0.5 mT to 0.9 mT in the IEC 60601-2-33 edition 4.0 standard.