Key result
Exposure to electromagnetic articulography fields caused 0 adverse effects in device operation among 12 adults with pacemakers or ICDs, suggesting safe use in this population.
Why the study?
Does exposure to electromagnetic articulography (EMA) magnetic fields cause interference or adverse effects in adults with pacemakers or ICDs?
Observational (n=12)
Does exposure to electromagnetic articulography (EMA) magnetic fields cause interference or adverse effects in adults with pacemakers or ICDs?
Electromagnetic articulography appears safe for use in patients with pacemakers and ICDs, though temporary telemetry interference may occur during testing.
Supports EMA safety in device patients from limited data; leaves open need for prospective validation before practice change.
PURPOSE: Electromagnetic articulography (EMA) uses a helmet to create alternating magnetic fields for tracking speech articulator movement. An important safety consideration is whether EMA magnetic fields interfere with the operation of speakers' pacemakers or implantable cardioverter-defibrillators (ICDs). In this investigation, individuals with pacemaker/ICD devices were exposed to EMA fields under controlled conditions while potential interference was examined. METHOD: Twelve adults with pacemaker/ICD devices from 3 major manufacturers were assessed for device function before, during, and after exposure to magnetic fields from a Carstens AG100 EMA system. Potential interference was probed, with EMA transmitters positioned at varying distances from the implantable devices and with the EMA system set at different operating strengths. RESULTS: No adverse affects in device operation were observed under any conditions. The only potential complication was temporary telemetry-link interference during device testing in some cases. CONCLUSION: The results suggest that EMA technology may be safely used with patients who have pacemakers and ICDs. However, the present findings do not rule out potential interference with other pacemaker/ICD manufacturers or with different articulography systems. Precautions are suggested for testing individuals with pacemaker/ICDs under EMA conditions.
No takes yet. Share an insight, caveat, or question.
Joglar et al. (2009) conducted an observational in Pacemakers and implantable cardioverter-defibrillators (ICDs) (n=12). Electromagnetic articulography (EMA) was evaluated on Adverse effects in device operation. Exposure to electromagnetic articulography fields caused 0 adverse effects in device operation among 12 adults with pacemakers or ICDs, suggesting safe use in this population.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: