In-silico VAD modeling identifies optimal speed changes up to ~300 rpm across simulated HF severities.
Why the study?
Cardiovascular system and mechanical circulatory support models can provide valuable insights into clinical procedures, motivating this demonstration of an in-silico CVS-VAD model for invasive hemodynamic ramp testing.
Can an in-silico cardiovascular-VAD model be used to perform hemodynamic ramp testing and optimize pump speed in virtual heart failure patients?
Can an in-silico cardiovascular-VAD model be used to perform hemodynamic ramp testing and optimize pump speed in virtual heart failure patients?
An open-source acausal cardiovascular-VAD model can successfully simulate hemodynamic ramp testing to determine optimal pump speeds for virtual patients with varying severities of heart failure.
BACKGROUND: While cardiovascular system and mechanical circulatory support devices are efficiently model the effect of disease and assistance, they can also lend valuable insights into clinical procedures. This study demonstrates the use of a CVS-VAD model for an invasive procedure; hemodynamic ramp testing, in-silico. METHODS: The CVS model is developed using validated models in literature, using Simscape™. An analytically derived pump model is calibrated for the HeartWare VAD. Dilated cardiomyopathy is used as an illustrative example of heart failure, and heart failure patients are created virtually by calibrating the model with requisite disease parameters obtained from published patient data. A clinically applied ramp study protocol is adopted whereby speed optimization is performed following clinically accepted hemodynamic normalization criteria. Trends in hemodynamic variables in response to pump speed increments are obtained. Optimal speed ranges are obtained for the three virtual patients based on target values of central venous pressure (CVP), pulmonary capillary wedge pressure (PCWP), cardiac output (CO), and mean arterial pressure (MAP) for hemodynamic stabilization. RESULTS: Appreciable speed changes in the mild case (300 rpm), slight changes in the moderate case (100 rpm), and no changes in the simulated severe case are possible. CONCLUSION: The study demonstrates a novel application of cardiovascular modeling using an open-source acausal model, which can be potentially beneficial for medical education and research.
No takes yet. Share an insight, caveat, or question.
Mehmood et al. (2023) studied Dilated cardiomyopathy / heart failure (n=3). In-silico hemodynamic ramp testing was evaluated on Optimal speed ranges based on target values of CVP, PCWP, CO, and MAP. In-silico hemodynamic ramp testing using a cardiovascular-VAD model identified optimal speed changes of 300 rpm in mild, 100 rpm in moderate, and 0 rpm in severe simulated heart failure cases.