Key result
Computer simulations demonstrated that the IVAS numeric model normalizes cardiac output in failing hearts, albeit with reduced pulse pressure, and allows for testing of control algorithms.
Answers are grounded in this paper and cite their sources.
Numeric modeling of a nonpulsatile rotary LVAD demonstrates its ability to normalize cardiac output in simulated failing hearts, providing a platform for testing control algorithms.
A numeric model consisting of a lump-parameter cardiovascular system (CVS) model and a model for the Cleveland Clinic Implantable Ventricular Assist System (IVAS), a nonpulsatile rotary pump designed to augment the failing left ventricle, are described in this paper. The purposes of this study were to 1) observe the hemodynamic interactions between CVS and IVAS under various physiologic and pathophysiologic conditions running at different speeds; and 2) allow testing and optimization of various IVAS control algorithms. An existing numeric model of CVS (24 coupled differential equations, representing all cardiac chambers and systemic and pulmonary vasculature) was modified to add the IVAS pump as an auxiliary chamber between the left ventricle and aorta with pressure-flow-speed characteristics derived from in vitro testing. Simulations were conducted for ventricles with normal and abnormal systolic and diastolic dysfunction at different exercise levels with the pump running at various speeds. Computer simulations show that 1) numeric modeling is useful for predicting hemodynamic response of CVS to IVAS in various circumstances; 2) IVAS results in normalization of cardiac output, especially in failing hearts, although with reduced pulse pressure; and 3) various control algorithms allowing adaptation of IVAS to physiologic demands of CVS could be developed based on the simulation study.
No takes yet. Share an insight, caveat, or question.
Zhou et al. (1999) studied Heart failure. Cleveland Clinic Implantable Ventricular Assist System (IVAS) numeric model was evaluated on Hemodynamic interactions and response of CVS to IVAS. Computer simulations demonstrated that the IVAS numeric model normalizes cardiac output in failing hearts, albeit with reduced pulse pressure, and allows for testing of control algorithms.
Synapse has enriched 10 related papers. Consider them for comparative context: