Key result
Implantable loop recorders detect unrecognized AF in 31% of hemodialysis patients over six months.
Why the study?
Atrial fibrillation is common in patients with kidney failure on hemodialysis, but its incidence and burden when monitored with an implantable loop recorder required determination.
Does implantable loop recorder monitoring detect previously unrecognized atrial fibrillation in patients with kidney failure on hemodialysis?
Cohort (n=66)
Does implantable loop recorder monitoring detect previously unrecognized atrial fibrillation in patients with kidney failure on hemodialysis?
Implantable loop recorder monitoring reveals a high incidence (31%) of previously unrecognized atrial fibrillation in patients with kidney failure on hemodialysis, suggesting it may be an underappreciated cause of stroke in this population.
Answers are grounded in this paper and cite their sources.
May support ILR screening for AF detection in hemodialysis; leaves open effects on stroke prevention.
IntroductionAtrial fibrillation (AF) is common in patients with kidney failure on hemodialysis (KF-HD). We determined both AF incidence and burden in patients with KF-HD using implantable loop recorder (ILR) monitoring.MethodsPatients with KF-HD were enrolled and received an ILR. In 6 monitoring months, the incidence of AF events lasting ≥6 minutes was captured. Demographic, clinical, and dialysis characteristics were collected, and associations with incident AF were estimated using negative binomial regression models and expressed as incidence rate ratios and 95% CIs.ResultsWe enrolled 66 patients with KF-HD (mean age = 56 years, 70% male); 59 (90%) were without previously diagnosed AF. AF lasting ≥6 minutes was detected in 18 of 59 subjects (31%) without previously diagnosed AF and in 5 of 7 subjects (71%) with a previous AF diagnosis. Among the 23 with detected AF, episodes were present on 16% of patient days. Although 14 of 23 patients (61%) had AF on <5% of monitored days, the average duration of AF episodes was <1 hour in 13 of 23 patients (52%). Among patients with AF ≥6 minutes, 19 of 23 (83%) had a CHA2DS2-VASc score ≥2. When investigating individual HD parameters, higher dialysate calcium (>2.5 vs. 2.5 mEq/l: incidence rate ratio = 0.62; 95% CI, 0.48–0.80) was associated with lower AF risk whereas higher dialysate bicarbonate concentrations (>35 vs. 35 mEq/l: incidence rate ratio = 3.18; 95% CI, 1.13–8.94) were associated with higher AF risk.ConclusionNew AF was detected in approximately one-third of patients with KF-HD. AF affects a substantial proportion of patient days and may be an underappreciated cause of stroke in KF-HD.
No takes yet. Share an insight, caveat, or question.
Koplan et al. (2021) conducted a cohort in Kidney failure on hemodialysis (n=66). Implantable loop recorder (ILR) monitoring was evaluated on Incidence of AF events lasting ≥6 minutes. Implantable loop recorder monitoring over 6 months detected previously unrecognized atrial fibrillation lasting ≥6 minutes in 31% of patients with kidney failure on hemodialysis.