Key result
In adenosine insensitive patients with ventricular arrhythmias from the outflow tract, irrigated catheter ablation significantly improved the ablation success rate to 94.1% compared to 73.3% with nonirrigated catheters.
Why the study?
Does nonirrigated catheter ablation compared to irrigated catheter ablation affect success and recurrence rates differently based on adenosine sensitivity in patients with ventricular outflow tract arrhythmias?
RCT (n=216)
Open-label
2:1 ratio to nonirrigated or irrigated catheter
No
Does nonirrigated catheter ablation compared to irrigated catheter ablation affect success and recurrence rates differently based on adenosine sensitivity in patients with ventricular outflow tract arrhythmias?
Adenosine insensitivity predicts lower success and higher recurrence rates with nonirrigated catheters for ventricular outflow tract arrhythmias, suggesting irrigated catheters should be preferred in these patients.
Answers are grounded in this paper and cite their sources.
Irrigated catheters should be preferred in adenosine-insensitive outflow tract VA; extends RCT evidence for adenosine-guided ablation technique selection.
BACKGROUND: A high ablation success rate for ventricular arrhythmia (VA) from outflow tract has been achieved, but some of them cannot be eliminated from endocardium. We investigated the association between adenosine sensitivity and ablation success/recurrence rates with a nonirrigated or an irrigated catheter. METHODS: According to adenosine test, all patients were divided into a sensitive group (S group) or an insensitive group (I group). The patients of each group were randomized into a nonirrigated catheter (NA) subgroup or an irrigated catheter (IA) subgroup with a 2:1 ratio. RESULTS: In S group of 122 patients (84 in NA subgroup), the ablation success rate was similar between two subgroups (94.7% vs. 90.5%, P > 0.05), but in I group of 94 patients (60 in NA subgroup), it was higher in IA subgroup (94.1%) than that in NA subgroup (73.3%, P < 0.05). The success rate using nonirrigated catheter was significantly higher in S group (90.5%) than that in I group (73.3%, P < 0.01), and the recurrence rate was lower in S group than that in I group (1.3%, vs. 13.6%, P < 0.05). On the contrary, the success rate and the recurrence rate using irrigated catheter were similar between S group and I group (94.7%, 94.1%, P > 0.05, vs. 2.8%, 6.3%, P > 0.05). CONCLUSIONS: Adenosine insensitivity is associated with a lower success rate and a higher recurrence rate for VA patients undergoing nonirrigated catheter ablation. Thus, irrigated catheters should be the first choice for VA ablation in adenosine insensitive patients.
No takes yet. Share an insight, caveat, or question.
Feng et al. (2015) conducted an RCT in Ventricular Premature Contractions/Tachycardia from the Ventricular Outflow Tract (n=216). Irrigated catheter ablation vs. Nonirrigated catheter ablation (4-mm-tip, 30-35 W) was evaluated on Ablation success rate in adenosine insensitive patients (p=<0.05). In adenosine insensitive patients with ventricular arrhythmias from the outflow tract, irrigated catheter ablation significantly improved the ablation success rate to 94.1% compared to 73.3% with nonirrigated catheters.
Synapse has enriched 10 related papers. Consider them for comparative context: