Key result
CD31-mimetic stent coating reduces neointima versus BMS in pig coronary arteries via rapid endothelialization.
Why the study?
Rapid endothelialization of bare metal stents is counterbalanced by inflammation-induced neointimal growth, while drug-eluting stents impair endothelialization and delay arterial healing.
Does CD31-mimetic stent coating improve endothelialization and reduce neointimal development compared to DES and BMS in a preclinical porcine model?
Does CD31-mimetic stent coating improve endothelialization and reduce neointimal development compared to DES and BMS in a preclinical porcine model?
CD31-mimetic stent coating promotes rapid endothelialization and reduces neointimal growth and thrombosis in a preclinical porcine model, suggesting improved biocompatibility over current BMS and DES.
Answers are grounded in this paper and cite their sources.
CD31-mimetic stent coatings may promote rapid endothelialization and limit neointimal growth in porcine models; leaves open translation to human outcomes.
AIMS: The rapid endothelialization of bare metal stents (BMS) is counterbalanced by inflammation-induced neointimal growth. Drug-eluting stents (DES) prevent leukocyte activation but impair endothelialization, delaying effective device integration into arterial walls. Previously, we have shown that engaging the vascular CD31 co-receptor is crucial for endothelial and leukocyte homeostasis and arterial healing. Furthermore, we have shown that a soluble synthetic peptide (known as P8RI) acts like a CD31 agonist. The aim of this study was to evaluate the effect of CD31-mimetic metal stent coating on the in vitro adherence of endothelial cells (ECs) and blood elements and the in vivo strut coverage and neointimal growth. METHODS AND RESULTS: We produced Cobalt Chromium discs and stents coated with a CD31-mimetic peptide through two procedures, plasma amination or dip-coating, both yielding comparable results. We found that CD31-mimetic discs significantly reduced the extent of primary human coronary artery EC and blood platelet/leukocyte activation in vitro. In vivo, CD31-mimetic stent properties were compared with those of DES and BMS by coronarography and microscopy at 7 and 28 days post-implantation in pig coronary arteries (n = 9 stents/group/timepoint). Seven days post-implantation, only CD31-mimetic struts were fully endothelialized with no activated platelets/leukocytes. At day 28, neointima development over CD31-mimetic stents was significantly reduced compared to BMS, appearing as a normal arterial media with the absence of thrombosis contrary to DES. CONCLUSION: CD31-mimetic coating favours vascular homeostasis and arterial wall healing, preventing in-stent stenosis and thrombosis. Hence, such coatings seem to improve the metal stent biocompatibility.
No takes yet. Share an insight, caveat, or question.
Diaz‐Rodriguez et al. (2021) studied Coronary stenting. CD31-mimetic metal stent coating vs. Bare metal stents (BMS) and drug-eluting stents (DES) was evaluated on In vitro adherence of endothelial cells and blood elements, and in vivo strut coverage and neointimal growth. CD31-mimetic stent coating achieved full endothelialization at 7 days and significantly reduced neointima development at 28 days compared to bare metal stents in pig coronary arteries.