Answers are grounded in this paper and cite their sources.
In vascular-space-occupancy (VASO)-MRI, cerebral blood volume (CBV)-weighted contrast is generated by applying a nonselective inversion pulse followed by imaging when blood water magnetization is zero. An uncertainty in VASO relates to the completeness of blood water nulling. Specifically, radio frequency (RF) coils produce a finite inversion volume, rendering the possibility of fresh, non-nulled blood. Here, VASO-functional MRI (fMRI) was performed for varying inversion volume and TR using body coil RF transmission. For thin inversion volume thickness (delta(tot) < 10 mm), VASO signal changes were positive (DeltaS/S = 2.1-2.6%). Signal changes were negative and varied in magnitude for intermediate inversion volumes (delta(tot) = 100-300 mm), yet did not differ significantly (P > 0.05) for delta(tot) > 300 mm. These data suggest that blood water is in steady state for delta(tot) > 300 mm. In this appropriate range, long-TR VASO data converged to a less negative value (DeltaS/S = -1.4% +/- 0.2%) than short-TR data (DeltaS/S = -2.2% +/- 0.2%), implying that cerebral blood flow or transit-state effects may influence VASO contrast at short TR.
No takes yet. Share an insight, caveat, or question.
Donahue et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: