Key result
The retrogated compressed sensing cine sequence (CSrg) significantly reduced mean scan time to 58.3 seconds compared to 485.4 seconds for the reference sequence, without compromising image quality or quantification accuracy.
Why the study?
Real-time compressed sensing cine provides reliable biventricular quantification but may alter image quality, prompting assessment of a retrogated segmented compressed sensing 2D cine sequence.
Does a retrogated segmented compressed sensing 2D cine sequence (CSrg) maintain accurate quantification of ventricular functional parameters and improve image quality compared to real-time compressed sensing (CSrt) and reference SSFP cine in adult patients undergoing rest CMR?
Cross-Sectional (n=30)
Blinded analysis
None
No
Does a retrogated segmented compressed sensing 2D cine sequence (CSrg) maintain accurate quantification of ventricular functional parameters and improve image quality compared to real-time compressed sensing (CSrt) and reference SSFP cine in adult patients undergoing rest CMR?
Retrogated compressed sensing cine MRI provides accurate biventricular quantification and high image quality in a significantly reduced scan time of about one minute.
Background and objective: Real-time compressed sensing cine (CSrt) provides reliable quantification for both ventricles but may alter image quality. The aim of this study was to assess image quality and the accuracy of left (LV) and right ventricular (RV) volumes, ejection fraction and mass quantifications based on a retrogated segmented compressed sensing 2D cine sequence (CSrg). Methods: Thirty patients were enrolled. Each patient underwent the reference retrogated segmented steady-state free precession cine sequence (SSFPref), the real-time CSrt cine and the segmented retrogated prototype CSrg sequence providing the same slices. Functional parameters quantification and image quality rating were performed on SSFPref and CSrg images sets. The edge sharpness, which is an estimate of the edge spread function, was assessed for the three sequences. Results: The mean scan time was: SSFPref = 485.4 ± 83.3 (SD) s (95% CI: 454.3–516.5) and CSrg = 58.3 ± 15.1 (SD) s (95% CI: 53.7–64.2) (p < 0.0001). CSrg subjective image quality score (median: 4; range: 2–4) was higher than the one provided by CSrt (median: 3; range: 2–4; p = 0.0008) and not different from SSFPref overall quality score (median: 4; range: 2–4; p = 0.31). CSrg provided similar LV and RV functional parameters to those assessed with SSFPref (p > 0.05). Edge sharpness was significantly better with CSrg (0.083 ± 0.013 (SD) pixel−1; 95% CI: 0.078–0.087) than with CSrt (0.070 ± 0.011 (SD) pixel−1; 95% CI: 0.066–0.074; p = 0.0004) and not different from the reference technique (0.075 ± 0.016 (SD) pixel−1; 95% CI: 0.069–0.081; p = 0.0516). Conclusions: CSrg cine provides in one minute an accurate quantification of LV and RV functional parameters without compromising subjective and objective image quality.
No takes yet. Share an insight, caveat, or question.
Longère et al. (2021) conducted a cross-sectional in Cardiac conditions requiring rest CMR (n=30). Retrogated segmented compressed sensing 2D cine sequence (CSrg) vs. Reference retrogated segmented steady-state free precession cine sequence (SSFPref) was evaluated on Mean scan time (p=<0.0001). The retrogated compressed sensing cine sequence (CSrg) significantly reduced mean scan time to 58.3 seconds compared to 485.4 seconds for the reference sequence, without compromising image quality or quantification accuracy.