Key result
Withings Scanwatch detects postoperative AF with ~69% sensitivity vs continuous telemetry.
Why the study?
Atrial fibrillation frequently complicates cardiac surgery but is often silent and challenging to detect, and smartwatches present a potential screening solution.
Does the Withings Scanwatch accurately detect postoperative atrial fibrillation compared to standard telemetry in patients after cardiac surgery?
Observational (n=260)
No
Does the Withings Scanwatch accurately detect postoperative atrial fibrillation compared to standard telemetry in patients after cardiac surgery?
The Withings Scanwatch has high specificity but moderate sensitivity for detecting postoperative atrial fibrillation, suggesting it can support but not replace conventional telemetry monitoring.
Answers are grounded in this paper and cite their sources.
May support adjunctive wearable monitoring post-cardiac surgery; leaves open prospective validation before replacing telemetry.
Background: Atrial fibrillation, the world's predominant cardiac arrhythmia, frequently emerges as a complication post-cardiac surgery, leading to serious outcomes like strokes, heart failures, and even death. Due to its often-silent nature, detecting it can be challenging. Smartwatches present a potential solution, offering screening that is more rigorous. Objective: This prospective observational study sought to assess the Withings Scanwatch's efficacy in identifying postoperative atrial fibrillation. Methods: After cardiac surgery, patients received a Withings Scanwatch. Over a span of 24 h, both the smartwatch's photoplethysmography sensor and standard telemetry kept track of any atrial fibrillation incidents. Results: At the end of the study, data from 260 patients was available for assessment. Atrial fibrillation was identified in 32 of these patients, either via telemetry or the smartwatch. Our data revealed a sensitivity of 69.0%, specificity of 98.7%, a positive predictive value of 87.0%, and a negative predictive value of 96.2%. Conclusions: This clinical study is the first to evaluate the photoplethysmography sensor of the Withings Scanwatch, and it shows that the Scanwatch has high a specificity and moderate sensitivity in detecting postoperative atrial fibrillation. Thus, Scanwatch may support the conventional screening for atrial fibrillation, and potentially reducing complications and costs of atrial fibrillation. Because of lower than expected sensitivity this technology cannot replace conventional monitoring in postoperative patients.
No takes yet. Share an insight, caveat, or question.
Schreier et al. (2026) conducted an observational in Postoperative atrial fibrillation (n=260). Withings Scanwatch vs. Telemetry monitoring was evaluated on Detection of postoperative atrial fibrillation (Sensitivity).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: