Key result
Smartwatch single-lead ECG algorithm detects AF with ~96% sensitivity compared to 12-lead ECG.
Why the study?
Atrial fibrillation can be difficult to diagnose when asymptomatic or paroxysmal, presenting an opportunity for wearable connected devices to improve detection.
Does an automatic algorithm using a single-lead smartwatch ECG accurately detect atrial fibrillation compared to a 12-lead reference ECG in adult patients?
Cross-Sectional (n=262)
Blinded assessors
Yes
Does an automatic algorithm using a single-lead smartwatch ECG accurately detect atrial fibrillation compared to a 12-lead reference ECG in adult patients?
An automatic algorithm using a single-lead ECG from an analog smartwatch demonstrated high sensitivity (96.3%) and specificity (100%) for detecting atrial fibrillation compared to a standard 12-lead ECG.
Answers are grounded in this paper and cite their sources.
Supports smartwatch AF detection feasibility in adults; leaves open prospective validation before clinical adoption.
BACKGROUND: Atrial fibrillation affects approximately 4% of the world's population and is one of the major causes of stroke, heart failure, sudden death, and cardiovascular morbidity. It can be difficult to diagnose when asymptomatic or in the paroxysmal stage, and its natural history is not well understood. New wearables and connected devices offer an opportunity to improve on this situation. OBJECTIVE: We aimed to validate an algorithm for the automatic detection of atrial fibrillation from a single-lead electrocardiogram taken with a smartwatch. METHODS: Eligible patients were recruited from 4 sites in Paris, France. Electrocardiograms (12-lead reference and single lead) were captured simultaneously. The electrocardiograms were reviewed by independent, blinded board-certified cardiologists. The sensitivity and specificity of the algorithm to detect atrial fibrillation and normal sinus rhythm were calculated. The quality of single-lead electrocardiograms (visibility and polarity of waves, interval durations, heart rate) was assessed in comparison with the gold standard (12-lead electrocardiogram). RESULTS: A total of 262 patients (atrial fibrillation: n=100, age: mean 74.3 years, SD 12.3; normal sinus rhythm: n=113, age: 61.8 years, SD 14.3; other arrhythmia: n=45, 66.9 years, SD 15.2; unreadable electrocardiograms: n=4) were included in the final analysis; 6.9% (18/262) were classified as Noise by the algorithm. Excluding other arrhythmias and Noise, the sensitivity for atrial fibrillation detection was 0.963 (95% CI lower bound 0.894), and the specificity was 1.000 (95% CI lower bound 0.967). Visibility and polarity accuracies were similar (1-lead electrocardiogram: P waves: 96.9%, QRS complexes: 99.2%, T waves: 91.2%; 12-lead electrocardiogram: P waves: 100%, QRS complexes: 98.8%, T waves: 99.5%). P-wave visibility accuracy was 99% (99/100) for patients with atrial fibrillation and 95.7% (155/162) for patients with normal sinus rhythm, other arrhythmias, and unreadable electrocardiograms. The absolute values of the mean differences in PR duration and QRS width were <3 ms, and more than 97% were <40 ms. The mean difference between the heart rates from the 1-lead electrocardiogram calculated by the algorithm and those calculated by cardiologists was 0.55 bpm. CONCLUSIONS: The algorithm demonstrated great diagnostic performance for atrial fibrillation detection. The smartwatch's single-lead electrocardiogram also demonstrated good quality for physician use in daily routine care. TRIAL REGISTRATION: ClinicalTrials.gov NCT04351386; http://clinicaltrials.gov/ct2/show/NCT04351386.
No takes yet. Share an insight, caveat, or question.
Campo et al. (2022) conducted a cross-sectional in Atrial fibrillation (n=262). Analog smartwatch single-lead electrocardiogram algorithm vs. 12-lead electrocardiogram was evaluated on Sensitivity and specificity of the algorithm to detect atrial fibrillation (Specificity 100%, 95% CI Sensitivity lower bound 0.894, Specificity lower bound 0.967). An algorithm for automatic detection of atrial fibrillation from a smartwatch single-lead ECG demonstrated a sensitivity of 0.963 and specificity of 1.000 compared to a 12-lead ECG.