Key result
Bioresorbable polymer everolimus-eluting stents resulted in a significantly lower in-device percentage diameter stenosis compared to bioresorbable scaffolds (12.5% vs 19.3%, p=0.01) at 6 to 8 months.
Why the study?
The performance of everolimus-eluting bioresorbable scaffolds compared to metallic drug-eluting stents with bioresorbable polymers remained unstudied.
Does bioresorbable polymer everolimus-eluting stents improve percentage diameter stenosis compared to everolimus-eluting bioresorbable scaffolds in patients undergoing coronary stenting for de novo lesions?
RCT (n=117)
Open-label
1:1 computer-generated sequence
Yes
Does bioresorbable polymer everolimus-eluting stents improve percentage diameter stenosis compared to everolimus-eluting bioresorbable scaffolds in patients undergoing coronary stenting for de novo lesions?
Absolute Event Rate: 12.5% vs 19.3%
p-value: p=0.01
Bioresorbable polymer everolimus-eluting stents demonstrated superior mid-term angiographic performance compared to fully bioresorbable scaffolds, though clinical event rates did not differ significantly up to 5 years.
BP-EES outperform BRS for mid-term angiographic results; confirms BRS inferiority and supports metallic stent preference in de novo lesions.
Background: The performance of an everolimus-eluting bioresorbable scaffold (BRS) was inferior to an everolimus-eluting metallic drug-eluting stent (DES) with permanent polymer, mainly due the mechanical features of BRS technology. The performance of BRS as compared to metallic DES with bioresorbable polymers remains unstudied. Methods: This prospective, randomized, multicenter, clinical trial enrolled patients who underwent coronary stenting for de novo coronary lesions. Patients were randomly assigned to bioresorbable polymer everolimus-eluting stents (BP-EES) or everolimus-eluting BRS. The primary endpoint was percentage diameter stenosis (in-device) at 6- to 8-month angiographic surveillance. The main secondary endpoint was the device-oriented composite endpoint (DOCE) of cardiac death/target vessel-myocardial infarction/target lesion revascularization assessed after 12 months and 5 years. Results: The trial was prematurely terminated after the enrollment of 117 of 230 patients (BP-EES, n = 60; BRS, n = 57) due to safety issues associated with BRS technology. The primary endpoint of in-device diameter stenosis at angiographic surveillance was 12.5 ± 7.7% with BP-EES versus 19.3 ± 16.5% with BRS (p = 0.01). The DOCE occurred in 5.0% in the BP-EES group versus 12.3% of patients in the BRS group (hazard ratio [HR] 2.48, 95% confidence interval [CI] 0.64–9.58, p = 0.19) after 12 months and in 11.7% in the BP-EES group versus 26.4% of patients in the BRS group (HR 2.38, 95% CI 0.97–5.84, p = 0.06) after 5 years. Conclusions: BP-EES showed superior mid-term angiographic performance compared with BRS. Clinical event rates did not differ significantly between the groups up to 5 years of follow-up. These results should be interpreted with caution in view of the premature discontinuation of the study.
No takes yet. Share an insight, caveat, or question.
Wiebe et al. (2024) conducted an RCT in De novo coronary lesions (n=117). Bioresorbable polymer everolimus-eluting stents (BP-EES) vs. Everolimus-eluting bioresorbable scaffolds (BRS) was evaluated on Percentage diameter stenosis (in-device) at 6- to 8-month angiographic surveillance (p=0.01). Bioresorbable polymer everolimus-eluting stents resulted in a significantly lower in-device percentage diameter stenosis compared to bioresorbable scaffolds (12.5% vs 19.3%, p=0.01) at 6 to 8 months.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: