Key result
Bioresorbable vascular scaffolds showed no significant difference in the device-oriented composite endpoint compared to everolimus-eluting stents at 1 year (OR 1.19; 95% CI 0.94-1.52; p=0.16).
Why the study?
Does the everolimus-eluting bioresorbable vascular scaffold (BVS) reduce the device-oriented composite endpoint compared to cobalt-chromium durable polymer everolimus-eluting stents (EES) in patients undergoing percutaneous coronary intervention?
Meta-Analysis (n=5,588)
Does the everolimus-eluting bioresorbable vascular scaffold (BVS) reduce the device-oriented composite endpoint compared to cobalt-chromium durable polymer everolimus-eluting stents (EES) in patients undergoing percutaneous coronary intervention?
Odds Ratio: 1.19 (95% CI 0.94–1.52)
Absolute Event Rate: 6% vs 4.9%
p-value: p=0.16
At 1-year follow-up, bioresorbable vascular scaffolds were associated with a higher risk of myocardial infarction and scaffold thrombosis compared to durable polymer everolimus-eluting stents, with no significant difference in the device-oriented composite endpoint.
BVS increases first-year events versus EES; confirms early hazard in pooled RCT data and supports avoiding routine use.
BACKGROUND: The Absorb bioresorbable vascular scaffold (BVS) was developed to address long-term safety issues of metallic drug-eluting stents. However, it may be associated with an increased event risk during the first year. METHODS: A systematic literature search was performed (in MEDLINE/PubMed, Cochrane CENTRAL, EMBASE, and scientific meeting abstracts) to identify studies that compared BVS and cobalt-chromium durable polymer everolimus-eluting stents (EES). For randomized clinical trials and non-randomized propensity score matched studies that reported 1-year outcome data, fixed/random-effects models were used to generate pooled estimates of outcomes, presented as odds ratios (OR) with 95%-confidence intervals (CI). RESULTS: The 1-year follow-up data of 6 trials with 5588 patients were analyzed. A device-oriented composite endpoint (DOCE - cardiac death, target vessel myocardial infarction (MI), or target lesion revascularization (TLR)) was reached by 308 BVS or EES patients (195/3253 vs. 113/2315). Meta-analysis showed that patients who received BVS had an increased risk of MI (4.3% vs. 2.3%; OR:1.63, 95%-CI: 1.18-2.25, p<0.01) and definite-or-probable scaffold thrombosis (1.3% vs. 0.6%; OR:2.10, 95%-CI: 1.13-3.87, p=0.02). However, there was no significant between-group difference in risk of DOCE (6.0% vs. 4.9%; OR:1.19, 95%-CI: 0.94-1.52, p=0.16), cardiac death (0.8% vs. 0.7%; OR:1.14, 95%-CI: 0.54-2.39, p=0.73), or TLR (2.5% vs. 2.5%; OR: 0.98, 95%-CI:0.69-1.40, p=0.92). CONCLUSIONS: During the first year of follow-up, patients treated with BVS had a higher incidence of MI and scaffold thrombosis. The risk of DOCE was not significantly different. As BVS may pay off later, future robust data on long-term clinical outcome will be of paramount importance.
No takes yet. Share an insight, caveat, or question.
Mukete et al. (2016) reported a meta-analysis. Absorb bioresorbable vascular scaffold (BVS) vs. Cobalt-chromium durable polymer everolimus-eluting stents (EES) was evaluated on Device-oriented composite endpoint (cardiac death, target vessel myocardial infarction, or target lesion revascularization) (OR 1.19, 95% CI 0.94-1.52, p=0.16). Bioresorbable vascular scaffolds showed no significant difference in the device-oriented composite endpoint compared to everolimus-eluting stents at 1 year (OR 1.19; 95% CI 0.94-1.52; p=0.16).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: