Key result
Transvenous AIV pacing maintains stable thresholds and preserved cardiac function at 2 years post-mechanical TVR.
Why the study?
For patients without a left bundle branch block, the optimal position of coronary sinus leads remains controversial.
Does left ventricular pacing through the anterior interventricular vein provide stable pacing and preserve left ventricular function in patients with mechanical tricuspid valve replacement?
Case Report (n=1)
Does left ventricular pacing through the anterior interventricular vein provide stable pacing and preserve left ventricular function in patients with mechanical tricuspid valve replacement?
Left ventricular pacing via the anterior interventricular vein is a feasible and reliable alternative for patients with mechanical tricuspid valves, providing stable pacing parameters and preserving left ventricular function.
Answers are grounded in this paper and cite their sources.
May offer alternative pacing site in mechanical tricuspid valve patients; leaves open confirmation in prospective series.
In patients with a mechanical tricuspid prosthetic valve, the transvenous position of a ventricular lead through the coronary sinus (CS) is a good alternative option to right ventricular or epicardial lead implantation. In cardiac resynchronization therapy, pacing the left ventricular lateral wall was considered the best site for the CS lead. However, for patients without a left bundle branch block, the best position of CS leads remains controversial. Here, we present a case of placing CS lead in the anterior interventricular vein. Measurements at implantation and 2 years' follow-up reported low pacing thresholds with good sensing thresholds. The electrocardiograph showed narrow QRS complexes (120 ms) and follow-up echocardiography at 2 years presented left ventricular ejection fraction 58.9%.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2024) conducted a case report in Mechanical tricuspid prosthetic valve requiring permanent pacing (n=1). Left ventricular pacing through the anterior interventricular vein was evaluated on Pacing and sensing thresholds, and left ventricular ejection fraction. Left ventricular pacing through the anterior interventricular vein in a patient with a mechanical tricuspid valve replacement was feasible, providing stable pacing thresholds and preserved cardiac function over 2 years.