Key result
Quadripolar LV lead in the coronary sinus avoids a fourth surgery and achieves reasonable pacing thresholds.
Why the study?
In the presence of a prosthetic tricuspid valve, permanent pacing is challenging due to right ventricle inaccessibility, and single LV leads in the coronary sinus have limitations regarding sensing and threshold.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Quadripolar LV-only single lead pacing in the coronary sinus offers a less invasive alternative to epicardial lead placement in patients with tricuspid mechanical valves and lead malfunction.
In the presence of prosthetic tricuspid valve, the inaccessibility to the right ventricle makes permanent pacing challenging. The placement of a left ventricle (LV) single lead in the coronary sinus (CS) is a well-accepted alternative, with some limitations regarding sensing and threshold. We describe a clinical case of a patient who had a previous LV only lead in the CS due to the presence of a prosthetic tricuspid valve and, after a surgical valvular intervention, presented with recurrent syncope episodes due to lead malfunction with lack of pacing capture and significant ventricular pauses. A quadripolar lead was chosen to be placed in the CS connected to a cardiac resynchronization therapy pacemaker device, programmed at biventricular VVI and using a specific manufacturer T-wave protection algorithm to prevent pacemaker-induced arrhythmias and to use the patient's own rhythm. This approach prevented a fourth surgical intervention to place an epicardial lead and resulted in reasonable LV sensing and pacing threshold. Learning objectives: This paper reports an alternative and atypical approach that could solve some of the limitations associated with ventricular pacing in patients with tricuspid prosthetic valves and multiple previous surgeries.
No takes yet. Share an insight, caveat, or question.
Grazina et al. (2022) conducted a case report in Prosthetic tricuspid valve and recurrent syncope due to lead malfunction (n=1). Quadripolar left ventricle single lead pacing in the coronary sinus was evaluated on Prevention of surgical intervention and achievement of reasonable LV sensing and pacing threshold. Placement of a quadripolar left ventricular lead in the coronary sinus prevented a fourth surgical intervention and achieved reasonable sensing and pacing thresholds.
Synapse has enriched 10 related papers. Consider them for comparative context: