Key result
Exposure to 50/60 Hz electric fields caused no failure in >99% of clinical configurations at ICNIRP public exposure levels, except for six pacemakers in unipolar mode with maximum sensitivity.
Why the study?
External extremely low-frequency electric and magnetic fields induce internal electric fields, creating uncertainty regarding whether and at what levels active implanted cardiac devices may be perturbed by 50 or 60 Hz electric fields.
Does exposure to 50 or 60 Hz electric fields perturb active implantable cardiac devices?
Does exposure to 50 or 60 Hz electric fields perturb active implantable cardiac devices?
Answers are grounded in this paper and cite their sources.
Implantable cardiac devices configured with nominal sensitivity in bipolar mode are resistant to 50/60 Hz electric fields exceeding low action levels.
Exposure to external extremely low-frequency (ELF) electric and magnetic fields induces the development of electric fields inside the human body, with their nature depending on multiple factors including the human body characteristics and frequency, amplitude, and wave shape of the field. The objective of this study was to determine whether active implanted cardiac devices may be perturbed by a 50 or 60 Hz electric field and at which level. A numerical method was used to design the experimental setup. Several configurations including disadvantageous scenarios, 11 implantable cardioverter-defibrillators, and 43 cardiac pacemakers were tested in vitro by an experimental bench test up to 100 kV/m at 50 Hz and 83 kV/m at 60 Hz. No failure was observed for ICNIRP public exposure levels for most configurations (in more than 99% of the clinical cases), except for six pacemakers tested in unipolar mode with maximum sensitivity and atrial sensing. The implants configured with a nominal sensitivity in the bipolar mode were found to be resistant to electric fields exceeding the low action levels, even for the highest action levels, as defined by the Directive 2013/35/EU. Bioelectromagnetics. 2020;41:136-147.
No takes yet. Share an insight, caveat, or question.
Gerçek et al. (2020) studied Active implantable cardiac devices (n=54). 50 or 60 Hz electric field exposure was evaluated on Device perturbation or failure. Exposure to 50/60 Hz electric fields caused no failure in >99% of clinical configurations at ICNIRP public exposure levels, except for six pacemakers in unipolar mode with maximum sensitivity.