Key result
Exposure to electric fields exceeding 5.1 kV/m from 400 kV power lines caused anomalous behavior in ICDs, whereas no effect on functioning was observed at exposures up to 0.9 kV/m.
Why the study?
Does exposure to electric and magnetic fields of 400 kV power lines cause malfunction in implantable cardioverter defibrillators?
Does exposure to electric and magnetic fields of 400 kV power lines cause malfunction in implantable cardioverter defibrillators?
ICDs may experience functional disturbances when exposed to electric fields exceeding 5 kV/m, such as those found in close proximity to 400 kV power lines.
Answers are grounded in this paper and cite their sources.
May warrant caution for ICD patients near high-voltage lines; animal data leaves open human translation.
BACKGROUND: Implantable cardioverter defibrillator (ICD) therapy has increased in Western countries. The aim of the study was to investigate the function of ICDs using a human-shaped phantom in electric and magnetic fields of 400 kV power lines. METHODS: The phantom was used in the following manner: isolated from the ground, earthed from a foot, or earthed from a hand. RESULTS: We performed 37 ICD tests using 10 different ICD devices. When the electric fields varied from 6.8 kV/m to 7.5 kV/m (humidity 70.5%) and the magnetic field was 2.0 μT, one of the ICDs tested recorded 258 ventricular beats/min when a simulated heart signal was applied to ICD electrodes. When the exposure was 5.1 kV/m, the same ICD had a similar disturbance; however, in a 0.9 kV/m field, it worked correctly. CONCLUSIONS: Consequently, no effect on ICDs functioning was observed up to 0.9 kV/m, while anomalous behavior in some conditions was observed when levels exceeded 5.1 kV/m; ICD malfunctioning seems possible within 11.5 m from 400 kV power lines or in conditions inducing exposures exceeding 5 kV/m. Further development of this research field is needed.
No takes yet. Share an insight, caveat, or question.
Korpinen et al. (2013) studied Implantable cardioverter defibrillator (ICD) therapy (n=10). Electric and magnetic fields of 400 kV power lines vs. Lower exposure fields (e.g., 0.9 kV/m) was evaluated on ICD function and anomalous behavior. Exposure to electric fields exceeding 5.1 kV/m from 400 kV power lines caused anomalous behavior in ICDs, whereas no effect on functioning was observed at exposures up to 0.9 kV/m.