Key result
In an 86-year-old man, over-torquing of the delivery sheath during left atrial appendage closure caused kinking and unintentional release of the Watchman FLX device, which self-deployed safely.
Why the study?
Device dislodgement during Watchman FLX implantation can occur and compromise device effectiveness.
Case Report (n=1)
No
This case highlights the risk of unintentional Watchman FLX device release due to delivery sheath kinking and emphasizes the need to replace kinked sheaths, while also cautioning against the unverified use of AI models like ChatGPT in medical writing.
Answers are grounded in this paper and cite their sources.
Over-torquing may kink the sheath and cause unintentional Watchman FLX release; this case leaves open systematic risk assessment.
Atrial fibrillation (AF) is a common heart rhythm disorder that increases the risk of stroke and other cardiovascular complications. The Watchman FLX device is a percutaneous device used for the closure of the left atrial appendage in patients with atrial fibrillation and prevents blood clots from forming, thereby decreasing the risk of stroke. However, device dislodgement during the procedure can occur and compromise the effectiveness of the device. In this case report, we present the experience of an 86-year-old man with a history of atrial fibrillation, coronary artery disease, hypertension, and recent gastrointestinal bleeding due to diverticulosis. During the procedure, the device became dislodged prior to the planned release from the shaft due to the difficult anatomy of the appendage, which caused the operator to over-torque and kink the delivery sheath. The kinked sheath prevented the transmission of torque to the appropriate region of the sheath and caused the device to unscrew prematurely. Fortunately, the device was self-deployed in a satisfactory position, and no further intervention was required. The patient did not experience any complications prior to discharge and follow-up echocardiography showed proper positioning of the device. This case highlights the importance of careful technique during the Watchman FLX procedure and the need to replace the delivery sheath if kinking is noted to prevent unintentional dislodgement of the device. In addition, the case was written with the aid of the artificial intelligence-powered language model ChatGPT and demonstrates its benefits as well as calls for caution while using it for medical writing.
No takes yet. Share an insight, caveat, or question.
Kattoor et al. (2023) conducted a case report in Atrial fibrillation (n=1). Watchman FLX device was evaluated on Procedural success and complications. In an 86-year-old man, over-torquing of the delivery sheath during left atrial appendage closure caused kinking and unintentional release of the Watchman FLX device, which self-deployed safely.