Key result
Magbot catheter matches RMT lesion efficacy pre-clinically while improving contact force and reducing steam pops.
Why the study?
A novel catheter (Magbot) is needed to enable seamless integration with the robotic magnetic navigation system and third-party mapping platforms.
Does the Magbot catheter improve contact force and lesion characteristics compared to the RMT catheter in an ex vivo porcine heart model?
Does the Magbot catheter improve contact force and lesion characteristics compared to the RMT catheter in an ex vivo porcine heart model?
Answers are grounded in this paper and cite their sources.
The novel Magbot catheter matches the RMT catheter in lesion formation efficacy while offering improved contact force stability and a better safety profile regarding steam pops in a preclinical model.
Abstract Background A novel catheter (Magbot) is needed to enable seamless integration with robotic magnetic navigation (RMN) system and third-party mapping platforms. Objective To assess the contact force (CF) and lesion characteristics of the Magbot catheter under RMN guidance, and to compare these parameters with the RMT catheter solely for the Carto system. Methods A total of 1200 CF measurements were collected for both the Magbot and RMT catheters when spanning different vector orientations and catheter extensions were obtained in a simulation study. Subsequently, RF energy was delivered to ex vivo porcine heart to create ablation lesions using three power settings (30 W/30 s, 40 W/20 s, 50 W/15 s). Results CF values exhibited highly dependent on both vector orientation and catheter extension. The Magbot catheter consistently yielded more favorable CF values across all anatomical configurations compared to the RMT catheter. While no statistically significant difference in lesion size was observed between the two catheters across all tested energy settings, a consistent trend favoring the Magbot was noted. Moreover, high-power ablation (50 W/15 s) generated lesions characterized by a significant increase in superficial width but a concomitant reduction in depth and overall volume. Notably, compared to the Magbot catheter, the RMT catheter was associated with a significantly higher incidence of steam pops, with risk escalating sharply at higher power levels. Conclusion The findings indicated that the Magbot catheter not only matched the RMT in core lesion formation efficacy but also offered distinct advantages in terms of CF stability and safety profile.
No takes yet. Share an insight, caveat, or question.
Luo et al. (2026) studied this question. Magbot catheter vs. RMT catheter was evaluated on Contact force and lesion characteristics. In a pre-clinical study of 1200 measurements, the Magbot catheter matched the RMT catheter in lesion formation efficacy while providing more favorable contact force and fewer steam pops.
Synapse has enriched 10 related papers. Consider them for comparative context: