Key result
The 2-element phased array coil imaged 92.6% of coronary segments compared to 83.7% and 76.3% for the surface and cardiac phased array coils, respectively, providing improved image quality and SNR.
Cross-Sectional (n=14)
Blinded observers
No
Answers are grounded in this paper and cite their sources.
A custom-made 2-element phased array coil provides an inexpensive and accessible method for performing high-resolution MR coronary angiography outside of advanced imaging centers.
Parallel imaging using sophisticated receiver coils has improved the clinical feasibility of magnetic resonance coronary angiography (MRCA). These techniques, however, are not readily available outside advanced imaging centers. Our custom-made 2-element phased array coil is readily and inexpensively assembled to address this limitation and enable the widespread application of MRCA.
No takes yet. Share an insight, caveat, or question.
Nguyen et al. (2009) conducted a cross-sectional in Healthy volunteers (MR Coronary Angiography) (n=14). 2-element phased array coil vs. 5-inch surface coil and cardiac phased array coil was evaluated on Anatomic coverage (percentage of coronary segments imaged) (p=0.002). The 2-element phased array coil imaged 92.6% of coronary segments compared to 83.7% and 76.3% for the surface and cardiac phased array coils, respectively, providing improved image quality and SNR.